EAGLE IRRADIATIONS EXTERNAL Analysis Students Workshop Shift Plans
GAMMA-ONLY NEEDLE NEEDI COULEX PLUNGER GENERAL
HIL102-128Cs HIL087-176Pt Nd148_2 176Pt HIL109-172Os HIL124-128Cs HIL061 Nd148 HIL106-119Cs HIL097-118Xe HIL099-134Sm April16 xdaq
  HIL099-134Sm, Page 13 of 14  ELOG logo
Entry  Sat Mar 11 22:52:48 2023, BS, Runs, , General, run_0055 

run control = 55

run data = /onlineNEDA_00/xData/hil099/run_0055_11-03-2023_00h33m38s/needleRU_i55_0055_0137.caendat

RU = 4.17 MB/s

LF =   3.11 MB/s

plunger:

plunger run = 1439

actual distance =  515 um

 

micrometer =   578.2 um

position = 282.3 um

voltage =  1.372 V

 

Beam dump: 0.75 nA 

Colimator :  1.1 enA

 

stop: 11.15 pm -> new plunger distance

 

Entry  Sat Mar 11 23:30:47 2023, AF, Runs, , General, run_0056 

run control = 56

run data =/onlineNEDA_00/xData/hil099/run_0056_11-03-2023_23h26m10s

RU = 7.3 MB/s

LF =   3.5 MB/s

plunger:

plunger run = 1445

actual distance =  262.7 um

 

micrometer =   266.2 um

position = 30 um

voltage =  3.05 V

 

Beam dump: 0.75 nA 

Colimator :  1.1 enA

 

stop:

Entry  Sun Mar 12 00:09:35 2023, AF, Problem Fixed, , Software, caen_shifted correction in spy & greware 
neda_psdcaen_positions.txt had weird values for three first detectors, especially neda_02.

It was:
NEDA_00_position_psdcaen    0
NEDA_01_position_psdcaen    0.22
NEDA_02_position_psdcaen    40

I (AF) make it:
NEDA_00_position_psdcaen    0.2
NEDA_01_position_psdcaen    0.2
NEDA_02_position_psdcaen    0.2
Entry  Sun Mar 12 03:05:45 2023, GJ, AF, Problems, , Network, no data flux visible in grafena, LF prints "write: Broken pipe" 
As in the title, see below.
The data are still being constantly written on the disks with usual rate (also visible on grafena).
Greware is also seeing data.

beeper output:
no new data about the growth for 37 s !!!
Noticed at Sun Mar 12 02:50:42 2023
and weirdo scream followed.


LF output:

write: Broken pipe
  0-LF_neda_2:        98 ev ( 107153)*30   Tot=26234475   2253/s
write: Broken pipe
  0-LF_eagle_1:      881 ev ( 120281)*31   Tot=295856141  24268/s
write: Broken pipe
write: Broken pipe
  0-LF_neda_3:         3 ev (   3909)*5980 Tot=375647125  30917/s
write: Broken pipe
  0-LF_neda_1:        92 ev (  92962)*40   Tot=32523205   2820/s
write: Broken pipe
write: Broken pipe
  0-LF_neda_0:       104 ev ( 117997)*33   Tot=38078515   3323/s
  0-LF_eagle_0:      720 ev (  63994)*31   Tot=268339966  22243/s
write: Broken pipe
write: Broken pipe
write: Broken pipe
write: Broken pipe
write: Broken pipe
write: Broken pipe
  0-LF_neda_2:        72 ev (  87376)*31   Tot=26236761   2274/s
write: Broken pipe
  0-LF_eagle_1:      714 ev ( 104681)*35   Tot=295880265  23988/s
write: Broken pipe
write: Broken pipe
  0-LF_neda_3:         9 ev (   3988)*5870 Tot=375677518  30545/s
write: Broken pipe
  0-LF_neda_1:        93 ev (  98610)*34   Tot=32525919   2698/s
Entry  Sun Mar 12 03:23:05 2023, AF, Runs, , General, run_0057 

run control = 57

run data =/onlineNEDA_00/xData/hil099/run_0057_12-03-2023_03h10m10s/

RU = 8.1 MB/s

LF =   3.8 MB/s

plunger:

plunger run = 1484

actual distance =  262.7 um

 

micrometer =   266.9 um

position = 30 um

voltage =  3.306 V

 

Beam dump: 0.75 nA 

Colimator :  1.1 enA

 

stop:

Entry  Sun Mar 12 04:20:19 2023, OST, Runs, , General, Run 57 stopped, Run 58 started  
 
Entry  Sun Mar 12 07:01:31 2023, OST, Runs, , General,  

run control = 58

run data =xdaq 2254808224 Mar 12 06:59 needleRU_i58_0058_0017.caendat

RU = 8.10 MB/s

LF =   3.79 MB/s

plunger:

plunger run = 1520

actual distance =  262.7 um

 

micrometer =   266.7 um

position = 30 um

voltage =  3.306 V

 

Beam dump: 1 nA 

Colimator :  0.5 nA

Entry  Sun Mar 12 08:03:08 2023, OST, Runs, , General,  

run control = 58

run data = xdaq 3303602776 Mar 12 08:01 needleRU_i58_0058_0024.caendat

RU = 7.88 MB/s

LF =   3.70 MB/s

plunger:

plunger run = 1530

actual distance =  262.6 um

 

micrometer =   266.8 um

position = 29.9 um

voltage =  3.308 V

 

Beam dump: 1 nA 

Colimator :  0.5 nA

Entry  Sun Mar 12 09:15:00 2023, EG, Other, , , Array_monitors mat refresh synchronization 

Beam stable, counting rate within limits, plunger distance within the given tollerance, gamma spectra ok, everything looks good but....

(yes there must be some "but"), refreshing of EAGLE_array_monitor.mat and NEDA_array_monitor.mat are out of sync. Mostly pure red colour is dysplayed. Not very usefull and (I suppose) missleading information.


 

Entry  Sun Mar 12 13:20:00 2023, EG, Problems, , , No Data 

For a split second a red "NO DATA" notification appeared in the top right corner of Counting rates dashboard.

It looks like it did not change anything in data collection.


 

Entry  Sun Mar 12 17:03:12 2023, OST, Runs, , General,  

run control = 58

run data =xdaq 1111949504 Mar 12 17:01 needleRU_i58_0058_0083.caendat

RU = 7.14 MB/s

LF =   3.46 MB/s

plunger:

plunger run = 1620

actual distance =  262.5 um

 

micrometer =   266.7 um

position = 29.8 um

voltage =  3.31 V

 

Beam dump: 0.7 nA 

Colimator :  0.5 nA

Entry  Sun Mar 12 18:50:37 2023, OST, Runs, , General,  

run control = 58

run data =xdaq  471656980 Mar 12 18:47 needleRU_i58_0058_0095.caendat

RU = 7.35 MB/s

LF =   3.50 MB/s

plunger:

plunger run = 1638

actual distance =  262.6 um

 

micrometer =   266.9 um

position = 29.8 um

voltage =  3.308 V

 

Beam dump: 0.7 nA 

Colimator :  0.5 nA

Entry  Sun Mar 12 19:00:59 2023, OST, Runs, , General, Run 58 stopped and Run 59 started. 

18:54 Run 58 stopped and 18:55 Run 59 started.

BEAM STOPPED.

END of EXPERIMENT.

Run 59 collects radioactivity.

 

-rw-rw-r-- 1 xdaq xdaq 295178920 Mar 12 18:59 needleLF_i58_0059_0000.adf
-rw-rw-r-- 1 xdaq xdaq 802293316 Mar 12 18:59 needleRU_i59_0059_0000.caendat

Entry  Sun Mar 12 19:20:37 2023, MP, Runs, , General, run 60, radioactivity,  

Run 59 collected data using gg trigger radioactivty

Run 60 trigger changed fropm gg to single g.

Just in the beginning a few minutes gg data could have been collected.

Entry  Sun Mar 12 20:31:45 2023, GJ, Runs, , General, run_61 - target and stopper and beam dump dismounted. 
target and stopper and beam dump dismounted.
only the constructional elements of the plunger remained. 

pure background, g trigger only.

Michal says the"bg" spectrum did not change...
Entry  Sun Mar 12 20:55:08 2023, MP, Runs, , , runc62 152 for efficiency calibration 

started

source in the target position, as precisely as possible

 

Entry  Sun Mar 12 21:33:54 2023, GJ, Runs, , General, run 63 - 152 Eu during LN2 filling 
 
Entry  Sun Mar 12 22:52:36 2023, Bahadir Saygi, Other, , General, Plunger Run Files  134Sm_cont.zip134Sm.zip

Plunger run files for later to calculate error in the distance

Entry  Mon Mar 13 12:15:35 2023, MP, Other, , , measurement of radioactivity along the beam line 

At about 12:00, 13/03/2023 (about  17 hours after the beam was stopped), a measurement of activity of the beam line was performed using  FH 40G-L10 meter.

Radioctivity run 65 stopped before the measurement, radioactivity run  66 started after the measurement

Background level 120 ns/h

Target, stopper, the forward half of the target chamber - dismounted.

Approximate distance from the target position (upstream), approximate activity

10 cm    500 ns/h

25 cm    420 ns/h

35          350 ns/h

40          350 ns/h

50          200 ns/h

75          170 ns/h

 

 

Entry  Mon Mar 13 12:16:31 2023, MP, Other, , , run 66 

run 66 started, radiocativity (background)

 

ELOG V3.1.5-fc6679b