9:20 to 9:50 am beam current stayed 2.5 nA . After 9:50 am decrease 1.5 nA
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0000.caendat
/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleLF_i41_0042_0000.adf
RU = 6.01 MB/s
LF = 3.32 MB/s
Plunger ;
Plunger run = 105
Supposed Dist= 242.7 um
Micrometer = 246.1 um
Position = 9.9 um
Actual distance = 242.6 um
Voltage = 3.809 V
Beam dumb = 1.5 nA
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0004.caendat
RU = 5.69 MB/s
LF = 3.37 MB/s
Plunger run = 115
Micrometer = 246.4 um
Position = 10.3 um
Actual distance = 243 um
Voltage = 3.801 V
Data = /onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0010.caendat
RU = 5.46 MB/s
LF = 3.20 MB/s
Plunger run = 127
Micrometer = 245.9 um
Position = 10.4 um
Voltage = 3.8 V
Collimater=1.25 nA
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0019.caendat
RU = 5.12 MB/s
Plunger run = 145
Micrometer = 245.5 um
Position = 10 um
Actual distance = 242.7 um
Voltage = 3.812 V
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0026.caendat
RU = 5.24 MB/s
LF = 3.19 MB/s
Plunger run = 162
Voltage = 3.813 V
Beam dumb = 1 nA
Collimater=1 nA
16:30 beam increased 5 mb/s to 9 mb/s and stayed 1 hour 9 mb/s
17:10 beam fixed. stable
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0035.caendat
RU = 5.43 MB/s
LF = 3.10 MB/s
Plunger run = 179
Micrometer = 245 um
Position = 9.7 um
Actual distance = 242.4 um
Voltage = 3.822 V
Collimater=0.5 nA
9:08 local filter turned red. Both .caendat and LF .adf files still incrementing. Stopped run 44, started run 45, looks OK.
I started run 45, and local filter turn green everything is ok.
run control = 45
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0000.caendat
RU = 6.92 MB/s
LF = 3.71 MB/s
plunger:
plunger run = 340
actual distance = 282.7 um
supposed distance = 282.7 um
micrometer = 285.5 um
position = 50.1 um
voltage = 2.967 V
Beam dump = 1 nA
Collimater= 0.5 nA
after run 45 I reset all spectrums and number_Ge_15 dedector not taking data.
i check the fans_monitor_scaler (EAGLE_array_monitor) monitor and number 15 Ge, number ACS 5 never change the color just staying white there is a problem.
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0011.caendat
RU = 6.52 MB/s
LF = 3.57 MB/s
plunger run = 358
actual distance = 282.8 um
micrometer = 285.3 um
voltage = 2.965 V
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0019.caendat
RU = 6.42 MB/s
LF = 3.65 MB/s
plunger run = 373
actual distance = 283 um
micrometer = 285.2 um
position = 50.3 um
voltage = 2.962 V
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0025.caendat
RU = 6.54 MB/s
LF = 3.60 MB/s
plunger run = 384
voltage = 2.963 V
14:30 pm to 15:00 run 45 crashed and stopped
VME crate with CAEN digitizers was rebooted.
Run 46 started, number 15_Ge and number 5 anticompton dedector working now.
run control = 46
run data =/onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0004.caendat
RU = 6.51 MB/s
LF = 3.44 MB/s
plunger run = 405
micrometer = 285.9 um
position = 50 um
voltage = 2.966 V
Beam dump = 0.7 nA
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0064.caendat
RU = 5.75 MB/s
LF = 3.50 MB/s
plunger run = 676
actual distance = 382.8 um
supposed distance = 382.7 um
micrometer = 384.9 um
position = 150.4 um
voltage = 2.146 V
Beam dump 0.7