EAGLE IRRADIATIONS EXTERNAL Analysis Students Workshop Shift Plans
GAMMA-ONLY NEEDLE NEEDI COULEX PLUNGER GENERAL
HIL102-128Cs HIL087-176Pt Nd148_2 176Pt HIL109-172Os HIL124-128Cs HIL061 Nd148 HIL106-119Cs HIL097-118Xe HIL099-134Sm April16 xdaq
  HIL099-134Sm, Page 5 of 14  ELOG logo
Entry  Thu Mar 9 12:06:30 2023, OST, Runs, , General,  

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0070.caendat

RU =  6.89 MB/s

LF =   3.42 MB/s

plunger:

plunger run = 847

actual distance =  532.2 um

supposed distance = 532.7 um

micrometer =   533.6 um

position =  299.5 um

voltage =  1.659 V

 

Beam dump: 1 nA

Colimator :  0.5 enA

Entry  Thu Mar 9 10:05:30 2023, OST, Runs, , General,  

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0058.caendat

RU =  7.48 MB/s

LF =   3.69 MB/s

plunger:

plunger run = 827

actual distance =  532.4 um

supposed distance = 532.7 um

micrometer =   533.7 um

position =  300.4 um

voltage =  1.658 V

 

Beam dump: 1 nA

Colimator :  0.5 enA

Entry  Thu Mar 9 08:54:13 2023, OST, Runs, , General,  

run control = 50

run data =/onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0051.caendat

RU =  6.86 MB/s

LF =   3.42 MB/s

plunger:

plunger run = 815

actual distance =  533.6 um

supposed distance = 532.7 um

micrometer =   533.3 um

position =  299.9 um

voltage =  1.658 V

 

Beam dump: 1 nA

Colimator :  0.5 enA

Entry  Thu Mar 9 07:28:59 2023, WP, Runs, , General, General 

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0042.caendat

 

 

 

RU =  7.20 MB/s

LF =   3.54 MB/s

plunger:

plunger run = 801

actual distance =  533.2 um

supposed distance = 532.7 um

micrometer =   533.6 um

position =  300.7 um

voltage =  1.66 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

Entry  Thu Mar 9 06:24:10 2023, WP, Runs, , General, General 

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0035.caendat

 

 

RU =  7.39 MB/s

LF =   3.60 MB/s

plunger:

plunger run = 790

actual distance =  532.8 um

supposed distance = 532.7 um

micrometer =   533.6 um

position =  300.1 um

voltage =  1.657 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

Entry  Thu Mar 9 05:19:33 2023, WP, Runs, , General, General 

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0028.caendat

 

 

RU =  7.15 MB/s

LF =   3.68 MB/s

plunger:

plunger run = 779

actual distance =  533.1 um

supposed distance = 532.7 um

micrometer =   533.4 um

position =  299.6 um

voltage =  1.66 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

Entry  Thu Mar 9 04:17:21 2023, WP, Runs, , General, General 

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0022.caendat

 

 

RU =  7.48 MB/s

LF =   3.59 MB/s

plunger:

plunger run = 769

actual distance =  531.1 um

supposed distance = 532.7 um

micrometer =   533.7 um

position =  300.5 um

voltage =  1.655 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

Entry  Thu Mar 9 03:14:39 2023, WP, Runs, , General, General 

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0015.caendat

 

 

RU =  6.91 MB/s

LF =   3.48 MB/s

plunger:

plunger run = 758

actual distance =  533.5 um

supposed distance = 532.7 um

micrometer =   533.1 um

position =  300.9 um

voltage =  1.658 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

Entry  Thu Mar 9 02:14:09 2023, WP, Runs, , General, General 

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0008.caendat

 

 

RU =  7.24 MB/s

LF =   3.64 MB/s

plunger:

plunger run = 748

actual distance =  532.1 um

supposed distance = 532.7 um

micrometer =   532.9 um

position =  299.4 um

voltage =  1.66 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

Entry  Thu Mar 9 01:41:50 2023, GJ, Problem Fixed, , Hardware, TS zeroing - cables to S-IN from VME controller - between runs 49 and 50 
Since we were observing jumps of 16 ns in timings of all channels of whole boards, the assumption was that the 
signals from logical FIFO arrive at different times to S-IN inputs to zero the TS counters.
6 originally placed cables of quite the same length were replaced with 6 cables of declared 3-ns length, same 
producer, borrowed from NW's equipment.
Entry  Thu Mar 9 01:07:48 2023, WP, Runs, , General, New run started 50 (distance change) 

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0000.caendat

 

RU =  7.43MB/s

LF =   3.61 MB/s

plunger:

plunger run = 736

actual distance =  533 um

supposed distance = 532.7 um

micrometer =   533.5 um

position =  300 um

voltage =  1.657 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

Entry  Wed Mar 8 22:36:30 2023, OST, Runs, , General,  

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0082.caendat

RU =  6.06 MB/s

LF =   3.44 MB/s

plunger:

plunger run = 709

actual distance =  382.9 um

supposed distance = 382.7 um

micrometer =   384.7 um

position =  150.2 um

voltage =  2.145 V

 

Beam dump: 1 nA

Colimator :  0.5 nA

Entry  Wed Mar 8 20:46:16 2023, OST, Runs, , General,  

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0074.caendat

RU =  6.35 MB/s

LF =   3.47 MB/s

plunger:

plunger run = 693

actual distance =  382.5 um

supposed distance = 382.7 um

micrometer =   384.1 um

position =  149.8 um

voltage =  2.148 V

 

Beam dump: 1 nA

Colimator :  0.5 nA

Entry  Wed Mar 8 19:01:49 2023, OST, Runs, , General, beam current was very low and beam increased after this entry 

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0064.caendat

RU =  5.75 MB/s

LF =   3.50 MB/s

plunger:

plunger run = 676

actual distance =  382.8 um

supposed distance = 382.7 um

micrometer =   384.9 um

position =  150.4 um

voltage =  2.146 V

 

Beam dump 0.7

Entry  Wed Mar 8 18:02:43 2023, BS, Runs, , General, Rates Screenshot_2023-03-08_18-01-49.png
 
Entry  Wed Mar 8 17:59:38 2023, BS, Runs, , General, General 

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0058.caendat

RU =  9.83 MB/s

LF =   4.87 MB/s

plunger:

plunger run = 665

actual distance =  382.7 um

supposed distance = 382.7 um

micrometer =   384.4 um

position =  150.0 um

voltage =  2.148 V

 

Beam dump 1,25 ~ 1.5  nA

Entry  Wed Mar 8 16:53:07 2023, BS, Runs, , General, Rates Screenshot_2023-03-08_16-52-03.pngScreenshot_2023-03-08_16-51-42.png
 
Entry  Wed Mar 8 16:51:13 2023, BS, Runs, , General,  

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0050.caendat

RU =  6.75 MB/s

LF =   3.44 MB/s

plunger:

plunger run = 654

actual distance =  382.7 um

supposed distance = 382.7 um

micrometer =   384.4 um

position =  150.0 um

voltage =  2.147 V

 

Beam dump 1 nA

Entry  Wed Mar 8 15:57:53 2023, IK, AM, Problems, , General, Beam issue 

no beam in 15:52 - 16:21 due to power supply problem

Entry  Wed Mar 8 14:47:34 2023, AM, Runs, , Beam, beam unstable 

two drops in beam intesitivity at 14:33, 14:37 and 14:51

ELOG V3.1.5-fc6679b