There was a problem with the beam from 20:19 to 20:30.
Now it is fine.
run control = 46
run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0022.caendat
RU = 6.22 MB/s
LF = 3.34 MB/s
plunger:
plunger run = 439
actual distance = 282.5um
supposed distance = 282.7 um
micrometer = 284.9 um
position = 49.8 um
voltage = 2.97 V
Beam dump = 0.7 nA
Collimater= 0.5 nA
run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0013.caendat
RU = 6.10 MB/s
LF = 3.26 MB/s
plunger run = 422
actual distance = 282.6um
position = 49.9 um
voltage = 2.968 V
run data =/onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0004.caendat
RU = 6.51 MB/s
LF = 3.44 MB/s
plunger run = 405
actual distance = 283 um
micrometer = 285.9 um
position = 50 um
voltage = 2.966 V
VME crate with CAEN digitizers was rebooted.
Run 46 started, number 15_Ge and number 5 anticompton dedector working now.
14:30 pm to 15:00 run 45 crashed and stopped
/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0027.caendat
/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleLF_i44_0045_0010.adf
run control = 45
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0025.caendat
RU = 6.54 MB/s
LF = 3.60 MB/s
plunger run = 384
micrometer = 285.2 um
position = 50.3 um
voltage = 2.963 V
Beam dump = 1 nA
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0019.caendat
RU = 6.42 MB/s
LF = 3.65 MB/s
plunger run = 373
voltage = 2.962 V
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0011.caendat
RU = 6.52 MB/s
LF = 3.57 MB/s
plunger run = 358
actual distance = 282.8 um
micrometer = 285.3 um
position = 50.1 um
voltage = 2.965 V
after run 45 I reset all spectrums and number_Ge_15 dedector not taking data.
i check the fans_monitor_scaler (EAGLE_array_monitor) monitor and number 15 Ge, number ACS 5 never change the color just staying white there is a problem.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0000.caendat
RU = 6.92 MB/s
LF = 3.71 MB/s
plunger run = 340
actual distance = 282.7 um
micrometer = 285.5 um
voltage = 2.967 V
I started run 45, and local filter turn green everything is ok.
9:08 local filter turned red. Both .caendat and LF .adf files still incrementing. Stopped run 44, started run 45, looks OK.
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0023.caendat
RU = 6.57 MB/s
LF = 3.45 MB/s
plunger run = 318
actual distance = 282.6 um
micrometer = 285.1 um
position = 50.0 um
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0017.caendat
RU =6.95 MB/s
LF = 3.54MB/s
plunger run = 308
Beam dump = 0.7-1 nA