| ID |
Date |
Author |
Type |
test |
Category |
Subject |
|
120
|
Tue Mar 7 02:02:18 2023 |
JD, DW | Runs | | General | |
run control = 43
run data = /onlineNEDA_00/xData/hil099/run_0043_07-03-2023_00h50m41s/needleRU_i43_0043_0006.caendat
RU = 6.27 MB/s
LF = 3.41 MB/s
plunger:
plunger run = 265
actual distance = 282.9 um
supposed distance = 282.7um
micrometer = 285.2 um
position = 50.2 um
voltage = 2.963V
Beam dump = 1 nA
Collimater= 0.5 nA
|
|
121
|
Tue Mar 7 03:21:33 2023 |
JD, DW | Problem Fixed | | General | daq crash |
daq crashed at 02:59:00, resolved at 03:16:00
|
|
122
|
Tue Mar 7 03:26:54 2023 |
JD, DW | Runs | | General | Run 44 started |
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0000.caendat
RU = 6.46 MB/s
LF = 3.39 MB/s
plunger:
plunger run = 279
actual distance = 282.7 um
supposed distance = 282.7 um
micrometer = 285.1 um
position = 49.9 um
voltage = 2.966 V
Beam dump = 0.7-1 nA
Collimater= 0.5 nA |
|
123
|
Tue Mar 7 04:16:28 2023 |
JD, DW | Runs | | General | |
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0005.caendat
RU = 6.49 MB/s
LF = 3.42 MB/s
plunger:
plunger run = 288
actual distance = 282.6um
supposed distance = 282.7um
micrometer = 285.1um
position = 49.9 um
voltage = 2.967V
Beam dump = 0.7-1nA
Collimater= 0.5nA
ELOG V3.0.0-21dc563 |
|
124
|
Tue Mar 7 05:19:55 2023 |
JD, DW | Runs | | General | |
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0011.caendat
RU = 6.48 MB/s
LF = 3.42 MB/s
plunger:
plunger run = 298
actual distance = 282.7 um
supposed distance = 282.7 um
micrometer = 285.0 um
position = 50.1 um
voltage = 2.967 V
Beam dump = 0.7-1 nA
Collimater= 0.5 nA |
|
125
|
Tue Mar 7 06:18:46 2023 |
JD, DW | Runs | | General | |
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0017.caendat
RU =6.95 MB/s
LF = 3.54MB/s
plunger:
plunger run = 308
actual distance = 282.6 um
supposed distance = 282.7 um
micrometer = 285.1 um
position = 49.9 um
voltage = 2.968 V
Beam dump = 0.7-1 nA
Collimater= 0.5 nA |
|
126
|
Tue Mar 7 07:21:04 2023 |
JD, DW | Runs | | General | |
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0023.caendat
RU = 6.57 MB/s
LF = 3.45 MB/s
plunger:
plunger run = 318
actual distance = 282.6 um
supposed distance = 282.7 um
micrometer = 285.1 um
position = 50.0 um
voltage = 2.967 V
Beam dump = 1 nA
Collimater= 0.5 nA |
|
127
|
Tue Mar 7 09:18:01 2023 |
OST | Problems | | Beam | run number 44 crashed and local filter turned red-but data saved |
9:08 local filter turned red. Both .caendat and LF .adf files still incrementing. Stopped run 44, started run 45, looks OK.
|
| Attachment 1: run_control_7mar2023_0910.png
|
|
|
128
|
Tue Mar 7 09:20:53 2023 |
OST | Problem Fixed | | Beam | run 45 start-local filter turn green |
I started run 45, and local filter turn green everything is ok. |
|
129
|
Tue Mar 7 09:24:53 2023 |
OST | Runs | | General | |
run control = 45
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0000.caendat
RU = 6.92 MB/s
LF = 3.71 MB/s
plunger:
plunger run = 340
actual distance = 282.7 um
supposed distance = 282.7 um
micrometer = 285.5 um
position = 50.1 um
voltage = 2.967 V
Beam dump = 1 nA
Collimater= 0.5 nA |
|
130
|
Tue Mar 7 09:41:44 2023 |
OST | Problems | | Software | number_Ge_15 dedector not taking data |
after run 45 I reset all spectrums and number_Ge_15 dedector not taking data.
i check the fans_monitor_scaler (EAGLE_array_monitor) monitor and number 15 Ge, number ACS 5 never change the color just staying white there is a problem.
|
|
131
|
Tue Mar 7 09:56:07 2023 |
OST | Problems | | mechanics | Beam stopped and operators controlling number 15_Ge dedector. |
|
|
132
|
Tue Mar 7 11:19:54 2023 |
OST | Runs | | General | |
run control = 45
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0011.caendat
RU = 6.52 MB/s
LF = 3.57 MB/s
plunger:
plunger run = 358
actual distance = 282.8 um
supposed distance = 282.7 um
micrometer = 285.3 um
position = 50.1 um
voltage = 2.965 V
Beam dump = 1 nA
Collimater= 0.5 nA |
|
133
|
Tue Mar 7 12:50:52 2023 |
OST | Runs | | General | |
run control = 45
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0019.caendat
RU = 6.42 MB/s
LF = 3.65 MB/s
plunger:
plunger run = 373
actual distance = 283 um
supposed distance = 282.7 um
micrometer = 285.2 um
position = 50.3 um
voltage = 2.962 V
Beam dump = 1 nA
Collimater= 0.5 nA |
|
134
|
Tue Mar 7 13:51:17 2023 |
OST | Runs | | General | |
run control = 45
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0025.caendat
RU = 6.54 MB/s
LF = 3.60 MB/s
plunger:
plunger run = 384
actual distance = 283 um
supposed distance = 282.7 um
micrometer = 285.2 um
position = 50.3 um
voltage = 2.963 V
Beam dump = 1 nA
Collimater= 0.5 nA |
|
135
|
Tue Mar 7 14:03:20 2023 |
BS | Runs | | General | Counting Rates |
|
| Attachment 1: Screenshot_2023-03-07_14-01-56.png
|
|
|
136
|
Tue Mar 7 14:07:33 2023 |
BS | Other | | General | Run Files Number |
/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0027.caendat
/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleLF_i44_0045_0010.adf |
|
137
|
Tue Mar 7 14:23:43 2023 |
BS | Other | | General | RU and BU Rates |
|
| Attachment 1: Screenshot_2023-03-07_14-22-35.png
|
|
|
138
|
Tue Mar 7 14:39:08 2023 |
BS | Other | | General | Runlist File ScreenShot Accesses on 07 MArch at 14:38 |
|
| Attachment 1: Screenshot_2023-03-07_14-37-51.png
|
|
|
139
|
Tue Mar 7 15:12:06 2023 |
OST | Problems | | Software | run 45 stopped |
14:30 pm to 15:00 run 45 crashed and stopped |