1. start a new run every 2 hours
2. go to the HV power supply of the DSSD. On the small blue screen you will see 4 channels. Write down tyhe numbers for ch0, ch2, ch3. Ch1 is off
3. collect one fast file and check with nearline if 60Ni is indeed the beam. In You should see 1333 keV line in the Compton clean spectrum. See attached
4. Check the DSSD spectrum. It should not change - to do this select "DSSD spectra each channel" histogram and project one channel, and compare with run_37 (the first one after restoring tandem)
5. Check the current (between 20 and 40 enA)
6. run nearline with all files - 37-current,excluding run 41
7. fill out the google spreadsheet - the hours should correspond to what is stored on the disk so put in the real time, not the daq time
8. in workspace 2 there are 2 terminals. Please after stopping the run do the rsync of raw data and nearline |