  12   Wed Mar 1 16:06:50 2023 BSRuns  Run 23 StoppedRun 23 Stopped due to some unexpected spikes
in the

voltage of collimator.
  
  13   Wed Mar 1 16:07:49 2023 BSRuns  Run 24 StartedRun 24 has been started but no beams at
the moment.

EAGLE is opened.
  
  21   Thu Mar 2 09:05:43 2023 DODProblems  Sulphur on aluminium fusion-evaporation reactionsUsing  the data which we've sorted
to date, I looked into the gamma-gamma coincidences
and have identified a number of nuclei e.g.
  
  27   Thu Mar 2 11:17:54 2023 BSRuns  Run 29 stoppedWe will change the trigger and we stopped
the run number 29
  
  37   Thu Mar 2 17:28:21 2023 BSRuns  Run 33 StartedRun 33 Started

Plunger Run : 48

Supposed Distance = 10 um
  
  40   Fri Mar 3 00:09:19 2023 MP, ATRuns  run 35Started after beam current was reduced
by a factor of 2

plunger run # 116
  
  56   Fri Mar 3 13:43:56 2023 IP. KWLRuns  511 annihilation and 279 Au lines in Ge ringsThe counting rate in 511 keV line for sum
of the Ge detectors placed at 101, 117 and
143 degrees (with approved ACS). The largest
 ge_rings101_117_143_acceptedACS_gold_line_279keV.pngge_rings101_117_143_acceptedACS.png 
  91   Sun Mar 5 22:32:56 2023 MPOther  change of the targetSee pictures in the attachment:

1) stopper seen from the back (upstream
view)
 6x 
  228   Sat Mar 11 09:29:48 2023 EGOther  current exeriment parameters as requested in the manual for shifters41_ge_single_theta 101,117,143 looks ok.
No peakdeformation of 511, other gammas FWHM
around half of that of511 - ok
  
  229   Sat Mar 11 10:46:37 2023 EGOtherBeam intensity  ch4 in total rate EGALE_) reached 2.4 kHz,
beam intensity has been reduced to fit the
required 2kHz.
  
  230   Sat Mar 11 11:39:55 2023 EGProblems  Buffers clearingat 11:34:50 sek on 11 march a spike in
counting rate of some Ge detectors is observed.
Buffers not cleared properly?
  
  231   Sat Mar 11 12:42:42 2023 EGOther  Beam intensity reductionAgain beam intensity gradually increased
reaching 2.4 kHz ontotal rate EAGLE_0 ch4. 
Beam intensity reduction requested. 1.8 kHz
  
  232   Sat Mar 11 14:15:55 2023 EGRuns  Beam intensity rapid increase 14:12 beam intensity instability. Rapid
increase of beam intensity - twice the previous
value. Returned to normal after 4 minutes.
  
  233   Sat Mar 11 14:42:28 2023 EGRuns  Beam instabilitiesLarge beam intensity instabilities from
14:10 till 14:40.  Beam intensity reduction
requested. CH4 on Total rate EAGLE_0 returned
  
  234   Sat Mar 11 15:16:31 2023 EGOther  Beam intensity reduced againAgain beam intensity to high. Reduced to
2 kHz on ch4 GE
  
  250   Sun Mar 12 09:15:00 2023 EGOther  Array_monitors mat refresh synchronizationBeam stable, counting rate within limits,
plunger distance within the given tollerance,
gamma spectra ok, everything looks good but....
  
  251   Sun Mar 12 13:20:00 2023 EGProblems  No DataFor a split second a red "NO DATA"
notification appeared in the top right corner
of Counting rates dashboard.
  
  257   Sun Mar 12 20:55:08 2023 MPRuns  runc62 152 for efficiency calibrationstarted

source in the target position,
as precisely as possible
  
  260   Mon Mar 13 12:15:35 2023 MPOther  measurement of radioactivity along the beam lineAt about 12:00, 13/03/2023 (about 
17 hours after the beam was stopped), a measurement
of activity of the beam line was performed
  
  261   Mon Mar 13 12:16:31 2023 MPOther  run 66run 66 started, radiocativity (background)

 
  