EAGLE
IRRADIATIONS
EXTERNAL
Analysis
Students Workshop
Shift Plans
GAMMA-ONLY
NEEDLE NEEDI
COULEX
PLUNGER
GENERAL
HIL102-128Cs
HIL087-176Pt
Nd148_2
176Pt
HIL109-172Os
HIL124-128Cs
HIL061
Nd148
HIL106-119Cs
HIL097-118Xe
HIL099-134Sm
April16
xdaq
HIL099-134Sm, Page 7 of 14
New
|
Find
|
Config
|
Logout
New entries since:
Thu Jan 1 01:00:00 1970
Full
| Summary |
Threaded
-- All entries --
Last day
Last 3 Days
Last week
Last month
Last 3 Months
Last 6 Months
Last Year
-- Type --
Runs
Problems
Problem Fixed
Other
264 Entries
Goto page
Previous
1
,
2
,
3
...
6
, 7,
8
...
12
,
13
,
14
Next
All
ID
Date
Author
Type
test
Category
Subject
Text
179
Thu Mar 9 05:19:33 2023
WP
Runs
General
General
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0028.caendat
180
Thu Mar 9 06:24:10 2023
WP
Runs
General
General
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0035.caendat
181
Thu Mar 9 07:28:59 2023
WP
Runs
General
General
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0042.caendat
63
Fri Mar 3 20:50:31 2023
IK, MP, GJ
Other
Data analysis
Ge spectra
Ge rates increasing, similar to what it
was 24h before. Asking to decrease beam intensity
by a factor of 2.
159
Wed Mar 8 10:27:26 2023
IK, BS
Other
General
Ge spectra
Count rates are still a bit high (~2.6kHz
for EAGLE det4). Ge spectra looks fine compared
to earlier ones.
164
Wed Mar 8 13:04:05 2023
IK
Runs
General
Ge spectra
Ge spectra after #161 - #162 beam tuning.
206
Fri Mar 10 16:14:19 2023
IP
Other
General
Ge - the rate is higher
From 13:00 the rate of the Ge [the chanel
4] is a bit higher (~2.15 kHz). And also
the rate of other detectors.
222
Fri Mar 10 22:40:00 2023
IP
Other
General
Ge - the rate is higher
the rates of detectors are still a bit
higher
135
Tue Mar 7 14:03:20 2023
BS
Runs
General
Counting Rates
39
Thu Mar 2 23:04:01 2023
GJ
Problem Fixed
Hardware
Connection of the plunger cables, grounding, beam read-out for the operators
Notes of the measurements with different conditions
are attached.
230
Sat Mar 11 11:39:55 2023
EG
Problems
Buffers clearing
at 11:34:50 sek on 11 march a spike in
counting rate of some Ge detectors is observed.
Buffers not cleared properly?
62
Fri Mar 3 19:41:27 2023
IK
Problems
Hardware
Beamdump readout issue Vol.2.
Issue described in #61 continues to
persist.
Beam stopped from 19:20 to 19:32,
61
Fri Mar 3 17:52:41 2023
IK
Problem Fixed
Hardware
Beamdump readout issue - solved
Current readout issues from beamdump. Beam
stopped from 17:46-17:52, run 36 ongoing.
MP & GJ checking the grounding
131
Tue Mar 7 09:56:07 2023
OST
Problems
mechanics
Beam stopped and operators controlling number 15_Ge dedector.
186
Thu Mar 9 13:37:33 2023
OST
Other
Beam
Beam stopped 13:37-13:49 for control again.
166
Wed Mar 8 15:57:53 2023
IK, AM
Problems
General
Beam issue
no beam in 15:52 - 16:21 due to power supply
problem
231
Sat Mar 11 12:42:42 2023
EG
Other
Beam intensity reduction
Again beam intensity gradually increased
reaching 2.4 kHz ontotal rate EAGLE_0 ch4.
Beam intensity reduction requested. 1.8 kHz
234
Sat Mar 11 15:16:31 2023
EG
Other
Beam intensity reduced again
Again beam intensity to high. Reduced to
2 kHz on ch4 GE
232
Sat Mar 11 14:15:55 2023
EG
Runs
Beam intensity rapid increase
14:12 beam intensity instability. Rapid
increase of beam intensity - twice the previous
value. Returned to normal after 4 minutes.
233
Sat Mar 11 14:42:28 2023
EG
Runs
Beam instabilities
Large beam intensity instabilities from
14:10 till 14:40. Beam intensity reduction
requested. CH4 on Total rate EAGLE_0 returned
Goto page
Previous
1
,
2
,
3
...
6
, 7,
8
...
12
,
13
,
14
Next
All