EAGLE IRRADIATIONS EXTERNAL Analysis Students Workshop Shift Plans
N-SI-85 exp22.41 N-SI-68 N-SI-69 N-SI-128 N-SI-137
  Coulex experiments, Page 5 of 7  ELOG logo
Entry  Sat Jun 24 03:04:01 2023, Iolanda, Runs, , General, new run 55 Capture_d’écran_du_2023-06-24_03-09-18.pngCapture_d’écran_du_2023-06-24_03-12-51.pngCapture_d’écran_du_2023-06-24_03-13-12.png
new run 54 started 03:04

beam current:  20 enA

data rate : 6.2 MB/s
    Reply  Sat Jun 24 05:00:14 2023, Iolanda, Runs, , General, new run 55 
run 55 stopped at 05:01

==> DAQ stopped                 at     
: 23-06-24 at 05h-27m-02s
Entry  Sat Jun 24 01:14:26 2023, Iolanda, Runs, , General, new run 54 Capture_d’écran_du_2023-06-24_01-08-03.pngCapture_d’écran_du_2023-06-24_01-14-34.pngCapture_d’écran_du_2023-06-24_01-15-05.png
new run 54 started 01:05

beam current:  26 enA

data rate : 6.6 MB/s
    Reply  Sat Jun 24 03:02:10 2023, Iolanda, Runs, , General, new run 54 
run 54 stopped at 03:03

==> /data_nuball2/N-SI-137/run_54.setup has been written                                                               
 
Entry  Fri Jun 23 23:21:22 2023, Iolanda, Runs, , General, new run 53 Capture_d’écran_du_2023-06-23_23-17-22.pngCapture_d’écran_du_2023-06-23_23-20-21.pngCapture_d’écran_du_2023-06-23_23-20-55.png
new run 53 started 23:05

beam current:  25 enA (retunned at the beginning of the run)

DSSD currents:
    Reply  Sat Jun 24 01:02:47 2023, Iolanda, Runs, , General, new run 53 
run 53 stopped at 01:03

==> shell name : ./.user_stop_trunkv3.sh
    User commentary      :
Entry  Fri Jun 23 21:47:26 2023, Giorgia, Runs, , General, new run 52 Capture_d’écran_du_2023-06-23_21-38-18.pngCapture_d’écran_du_2023-06-23_21-47-50.pngCapture_d’écran_du_2023-06-23_21-48-16.png
new run 52 started 21:36

beam current:  18 enA

DSSD currents:
Entry  Fri Jun 23 19:52:57 2023, Giorgia, Runs, , General, new run 51 Capture_d’écran_du_2023-06-23_19-40-04.pngCapture_d’écran_du_2023-06-23_19-53-24.pngCapture_d’écran_du_2023-06-23_19-54-05.png
new run 51 started 19:39

beam current:  15 enA

DSSD currents:
Entry  Fri Jun 23 19:34:28 2023, Giorgia, Runs, , General, new run 50 Capture_d’écran_du_2023-06-23_17-36-30.pngCapture_d’écran_du_2023-06-23_19-35-09.pngCapture_d’écran_du_2023-06-23_19-35-32.png
new run 50 started 17:35

beam current:  15 enA

DSSD currents:
Entry  Fri Jun 23 15:38:01 2023, Giorgia, Runs, , General, new run 49 Capture_d’écran_du_2023-06-23_15-36-44.pngCapture_d’écran_du_2023-06-23_15-38-01.pngCapture_d’écran_du_2023-06-23_15-39-29.png
new run 49 started 15:25

beam current: it was 9, we put it back at 21 enA

DSSD currents:
Entry  Fri Jun 23 13:18:56 2023, pjn, Runs, , General, new run 48 
new run 48 started 13:20

beam current: 17 enA

DSSD currents:
Entry  Fri Jun 23 11:41:55 2023, pjn, Runs, , General, new run 47 Capture_d’écran_du_2023-06-23_11-47-09.png
new run 46 started 11:42

beam current: 22enA

DSSD currents:
Entry  Fri Jun 23 09:18:38 2023, pjn, Runs, , General, new run 46 Capture_d’écran_du_2023-06-23_09-39-18.png
new run 46 started 9:18

beam current: 14enA

DSSD currents:
Entry  Fri Jun 23 07:19:57 2023, pjn, Runs, , General, new run 45  Capture_d’écran_du_2023-06-23_07-29-05.png
start 7:20

beam current 25 enA

DSSD currents:
Entry  Fri Jun 23 05:21:17 2023, KHK, Other, , General, instructions for shifters Capture_d’écran_du_2023-06-23_05-12-03.pngCapture_d’écran_du_2023-06-23_05-16-20.png
1. start a new run every 2 hours

2. go to the HV power supply of the DSSD. On the small blue screen you will see 4 channels. Write down tyhe numbers for ch0, ch2, ch3. Ch1 is
off
    Reply  Fri Jun 23 15:22:48 2023, Giorgia, Runs, , General, end run 48 Capture_d’écran_du_2023-06-23_15-22-51.png
stop run 15:24

 

==> /data_nuball2/N-SI-137/run_48.setup has been written         
    Reply  Fri Jun 23 07:04:35 2023, KHKm JSP, Runs, , General, end of run_43 
 




Quote:



after spark
    Reply  Fri Jun 23 12:59:53 2023, pjn, Runs, , General, end of run 47 
end of run to open the cave and refill the tank




Quote:



new run 46 started 11:42
    Reply  Fri Jun 23 09:16:54 2023, pjn, Runs, , General, end of run 45  
cloce to the end of run -beam intensity was very low - incresed but only o 14enA




Quote:



start 7:20
Entry  Tue Jun 20 21:22:21 2023, KHK, KWL, Problems, , Other, contamination in the chamber Capture_d’écran_du_2023-06-20_21-18-27.pngCapture_d’écran_du_2023-06-20_21-17-56.png
We see the the background lines coming probably from some activation inside the chamber as the target wheel is removed and we run only
with the alpha source at the moment.

The spectrum is the same means that something inside the chamber is activated, not the target 197Au used in the coulex run or neither it's
ELOG V3.1.5-fc6679b