ch0: 24.96V 10.87uA
ch2: 25.01V 6.26uA
ch3: 25.00V 4.86uA
File name : run Run number : 17 User commentary : 62Ni on 197Au, intensity 2 enA Start time : 2023-06-18 03:21:16.842674 ==> End of ./.user_start.sh
Run 17 stopped because accelerator discharged
We observed a decrease in the particle energies. We attached one channel comparing with the same but from run 5. We keep running.
beam current 20 enA
charge state 17
energy is yet to be checked (tuned to 210 MeV of 58Ni, Alain has to recalculate and change to 230 MeV of 62Ni)
data rate 7.6 MB/s
target position 6 - Au
==> DAQ started at : 23-06-16 at 09h-25m-01s ==> Shell name : ./.user_start.sh File name : test_beam_tune_58Ni_62Ni_2 Run number : 0 User commentary : hoppefully 62Ni Start time : 2023-06-16 09:24:29.048192 ==> End of ./.user_start.sh
Angle (°) Emin Emax (MeV) (MeV) 128 46.34 80.95 129 46.34 80.23 130 46.33 79.54 131 46.29 78.86 132 46.24 78.19 133 46.18 77.54 134 46.10 76.91 135 46.01 76.29 136 45.92 75.70 137 45.81 75.11 138 45.70 74.54 139 45.58 73.99 140 45.46 73.99 141 45.33 73.45 142 45.20 72.93 143 45.07 72.42 144 44.93 71.93 145 44.80 71.46 146 44.66 70.99 147 44.53 70.55 148 44.39 70.11 149 44.26 69.70 150 44.12 69.29 151 43.99 68.53 152 43.87 68.16 153 43.74 67.81 154 43.62 67.48
we reset the DAQ and did the module_restart with the card IP 192.168.3.78
no change
we restarted the HV DSSD - no change