EAGLE IRRADIATIONS EXTERNAL Analysis Students Workshop Shift Plans
GAMMA-ONLY NEEDLE NEEDI COULEX PLUNGER GENERAL
HIL102-128Cs HIL087-176Pt Nd148_2 176Pt HIL109-172Os HIL124-128Cs HIL061 Nd148 HIL106-119Cs HIL097-118Xe HIL099-134Sm April16 xdaq
  HIL099-134Sm, Page 4 of 14  ELOG logo
Entry  Wed Mar 1 12:51:33 2023, MP, Runs, , Beam, beam on the target 
Beam on the target:

target+stopper+beam dump: 25 enA   (1.7 pnA)    not stable
5 mm collimator:  0.8 enA 

data taking continued with run 0023
Entry  Fri Mar 3 10:09:20 2023, IP. KWL, Runs, , General, beam intensity decreased to 0.3 enA (at beam dump) on our request  Ge_beamInten_03enA_differentYscale.pngGe_beamInten_03enA_sameYscale.png

// we noticed yesterday that GE at 143 degrees counts more than those placed at 117 and 101. The 511 keV line (+ background) is much more pronounced for Ge at 143 degrees. To investigate this issue we decreased the beam intensity to 0.3 enA ( value given by operators at the beam dump ). 

Conclusion - decrease of the beam intensity does not explain above observations (se attached figs)

 

plunger run # 216

Beam dump 0.3 enA (given by operators) * 7 (following entry ID#39) = 2.1 enA

Collimator     enA

Run Data =/onlineNEDA_00/xData/hil099/run_0035_03-03-2023_00h04m26s/needleRU_i35_0035_0040.caendat

RU : 2.15 MB/s

LF : 5.60 MB/s

Supposed Distance = 10 um

position of the target = 10.6 um

micrometer = 8.8 um

Voltage = 7.8 V

Entry  Fri Mar 3 10:21:18 2023, IP. KWL, Runs, , General, beam intensity back to 1.25 enA at the beam dump 

we increased beam intensity to 1.25 enA (readout at the beam dump)

Entry  Fri Mar 3 13:54:41 2023, IP. KWL, Runs, , General, beam energy measurement 

beam stop due to the beam energy measurement

The beam energy: ~145 MeV

Entry  Wed Mar 1 00:14:28 2023, MP, Other, , Beam, beam energy 
Beam energy measured:  148.2 MeV
Mind that this may be an overestimate (by ~1-2 MeV)
Entry  Wed Mar 8 19:01:49 2023, OST, Runs, , General, beam current was very low and beam increased after this entry 

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0064.caendat

RU =  5.75 MB/s

LF =   3.50 MB/s

plunger:

plunger run = 676

actual distance =  382.8 um

supposed distance = 382.7 um

micrometer =   384.9 um

position =  150.4 um

voltage =  2.146 V

 

Beam dump 0.7

Entry  Tue Feb 28 23:10:36 2023, BS, Other, , , What we need to include for Plunger 

Plunger Run :

Micrometer :

Supposed Distance:

Position :

Entry  Thu Mar 9 01:41:50 2023, GJ, Problem Fixed, , Hardware, TS zeroing - cables to S-IN from VME controller - between runs 49 and 50 
Since we were observing jumps of 16 ns in timings of all channels of whole boards, the assumption was that the 
signals from logical FIFO arrive at different times to S-IN inputs to zero the TS counters.
6 originally placed cables of quite the same length were replaced with 6 cables of declared 3-ns length, same 
producer, borrowed from NW's equipment.
Entry  Thu Mar 2 09:05:43 2023, DOD, Problems, , , Sulphur on aluminium fusion-evaporation reactions 

Using  the data which we've sorted to date, I looked into the gamma-gamma coincidences and have identified a number of nuclei e.g. 50Cr, 52Mn, 55Fe which would be consistent with fusion-evaporation reactions involving the sulphur-32 projectiles and aluminium-27. Such reactions would yield a 59Cu comound system and these channels would be populated as a result of the evaporation of a combination of alphas, protons and neutrons.

These lines are intense and the conicdences are very clear and clean. I think this suggests that the beam is hitting some aluminium components within the plunger chamber. This could be the plunger vessel itself or the target frame perhaps.

There is another interesting observation: some of the identified transitions are emitted from levels which have sub-ps lifetimes. At this plunger distance, we might expect these lines to be fully Doppler shifted as they would decay in-flight well in advance of reaching the stopper. Yet, we observe these lines with their non-Doppler shifted energies suggesting that they are not being emitted in-flight but rather at rest. This might suggest that the 32S+27Al reactions are taking place within some thick piece of aluminium. Some of the lines show signs of having been shifted in energy to some extent which might indicate that not all of the ions are completely stopped. The observed lineshapes seem to be consistent with the short-lived states showing evidence of Doppler shift while the decay of the "longer-lived" states (~9 ps) are very sharp.

I haven't slept for about 24 hours so I may be talking absolute sh*te (famous Scottish word).

Entry  Tue Feb 28 23:08:35 2023, BS, Other, , , SHIFT LIST 

https://docs.google.com/spreadsheets/d/1e2homI9RCYrg369z8aaY3VMIPgapUPOv0zKISsl58vQ/edit?usp=sharing


The data from google sheet obtained 14:17 07/03/23

Initials Name Surname e-mail      
DOD Dave O'Donnell
david.odonnell@uws.ac.uk
     
GJ Grzegorz Jaworski
tatrofil@slcj.uw.edu.pl
           
ÖST Ömer Sefa Temiz
ostemiz@ankara.edu.tr
     
BS Bahadır Saygı
bsaygi@ankara.edu.tr
       
KWL Kasia Wrzosek-Lipska
wrzosek@slcj.uw.edu.pl
     
ZCH Zuzanna Chochulska
zuzanna.chochulska.stud@pw.edu.pl
     
MK Małgorzata Kurach
malgorzata.kurach.stud@pw.edu.pl
     
MP Michał Prędota
michal.predota.stud@pw.edu.pl
       
EG Ernest Grodner
grodner.ernest@gmail.com
     
BSK Bartosz Skura
bartosz.skura.stud@pw.edu.pl
     
FS Franziskus Spee
fspee@ikp.uni-koeln.de
     
MB Marcel Beckers
mbeckers@ikp.uni-koeln.de
     
AK Agnieszka Korgul
korgul@fuw.edu.pl
     
SP Serhii Panasenko
sergey.i.panasenko@gmail.com
     
AT Andrzej Tucholski
tucholski@slcj.uw.edu.pl
     
KS Krzysztof Solak
Krzysztof.solak@post.pl
     
AF Aleksandra Fijałkowska
aleksandra.fijalkowska@fuw.edu.pl
       
MGB M. Güray Budak
mgbudak@gmail.com
     
IK István Kuti kuti@atomki.hu      
JSP Justyna Samorajczyk-Pyśk
samorajczyk@slcj.uw.edu.pl
     
JD James Deary
James.Deary@uws.ac.uk
     
DW Dylan White
Dylan.White@uws.ac.uk
     
IP Iwona Piętka
iz.pietka@student.uw.edu.pl
     
WP Wiktor Poklepa
w.poklepa@student.uw.edu.pl
       
KSz Klara Szlęzak
k.szlezak@student.uw.edu.pl
       
       
       

 

Entry  Tue Mar 7 14:39:08 2023, BS, Other, , General, Runlist File ScreenShot Accesses on 07 MArch at 14:38 Screenshot_2023-03-07_14-37-51.png
 
Entry  Tue Mar 7 14:07:33 2023, BS, Other, , General, Run Files Number 

/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0027.caendat

/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleLF_i44_0045_0010.adf

Entry  Sun Mar 12 19:00:59 2023, OST, Runs, , General, Run 58 stopped and Run 59 started. 

18:54 Run 58 stopped and 18:55 Run 59 started.

BEAM STOPPED.

END of EXPERIMENT.

Run 59 collects radioactivity.

 

-rw-rw-r-- 1 xdaq xdaq 295178920 Mar 12 18:59 needleLF_i58_0059_0000.adf
-rw-rw-r-- 1 xdaq xdaq 802293316 Mar 12 18:59 needleRU_i59_0059_0000.caendat

Entry  Sun Mar 12 04:20:19 2023, OST, Runs, , General, Run 57 stopped, Run 58 started  
 
Entry  Sat Mar 11 00:34:51 2023, BS, Runs, , General, Run 55 Started New Distance 500 um 

Run 55 Started New Distance 500 um

Entry  Wed Mar 8 10:31:10 2023, IK, Runs, , General, Run 49 

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0008.caendat

RU =  10.6 MB/s

LF =   4.89MB/s

plunger:

plunger run = 591

actual distance =  383.0 um

supposed distance = 382.7 um

micrometer =   385.8 um

position =  150.3 um

voltage =  2.145 V

 

Beam dump 0.7 nA

 

 

 

 

 

 

Entry  Wed Mar 8 11:42:14 2023, IK, Problems, , General, Run 49 Screenshot_2023-03-08_11-38-47.pngScreenshot_2023-03-08_11-38-19.pngScreenshot_2023-03-08_11-37-37.png

Counting rates continued to increase (det4 > 3kHz). Beam current looked fine on the beam dump, but beam is fluctuating, and can reach 2nA on the collimator.  Asked for improved focusing.

Attached screenshots and Ge spectra from before changing anything.

 

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0020.caendat

RU =  13.3 MB/s

LF =   6.15MB/s

plunger:

plunger run = 603

actual distance =  382.7 um

supposed distance = 382.7 um

micrometer =   385.9 um

position =  150.0 um

voltage =  2.146 V

 

Beam dump 0.7 nA

Collimator ~1.5-2nA when the generator retunes itself (prob the beam gets defocused at those moments)

 

 

 

 

 

Entry  Wed Mar 8 13:02:38 2023, IK, Runs, , General, Run 49 

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0030.caendat

RU =  7.86 MB/s

LF =   3.87 MB/s

plunger:

plunger run = 616

actual distance =  382.7 um

supposed distance = 382.7 um

micrometer =   385.7 um

position =  150.0 um

voltage =  2.147 V

 

Beam dump 0.7 nA

 

 

 

 

 

 

Entry  Wed Mar 8 09:31:43 2023, IK, Runs, , General, Run 48 stopped, run 49 started 

Run 48 stopped at 9:29 to start a new run.

Run 49 is started okay at 9:31.

 

run control = 49

run data = /onlineNEDA_00/xData/hil099/run_0048_08-03-2023_01h08m57s/needleRU_i48_0048_0041.caendat

RU =  10.1 MB/s

LF =   4.68MB/s

plunger:

plunger run = 582

actual distance =  383.1 um

supposed distance = 382.7 um

micrometer =   385.8 um

position =  150.5 um

voltage =  2.145 V

 

Beam dump 0.7 nA

 

 

 

 

 

 

Entry  Wed Mar 8 04:24:40 2023, KS, KSz, Runs, , General, Run 48 

run control = 48

run data =  /onlineNEDA_00/xData/hil099/run_0048_08-03-2023_01h08m57s/needleRU_i48_0048_0017.caendat

RU =  5.75 MB/s

LF =   3.21 MB/s

plunger:

plunger run = 530

actual distance =  382,7 um

supposed distance = 382.7 um

micrometer =   384.3 um

position =  150.3 um

voltage =  2.146 V

Beam dump =  1.5 nA

Collimater=  0.7 nA (?)

ELOG V3.1.5-fc6679b