ID Date Author Type test Category Subjectdown
  179   Thu Mar 9 05:19:33 2023 WPRuns GeneralGeneral

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0028.caendat

 

 

RU =  7.15 MB/s

LF =   3.68 MB/s

plunger:

plunger run = 779

actual distance =  533.1 um

supposed distance = 532.7 um

micrometer =   533.4 um

position =  299.6 um

voltage =  1.66 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

  180   Thu Mar 9 06:24:10 2023 WPRuns GeneralGeneral

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0035.caendat

 

 

RU =  7.39 MB/s

LF =   3.60 MB/s

plunger:

plunger run = 790

actual distance =  532.8 um

supposed distance = 532.7 um

micrometer =   533.6 um

position =  300.1 um

voltage =  1.657 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

  181   Thu Mar 9 07:28:59 2023 WPRuns GeneralGeneral

run control = 50

run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0042.caendat

 

 

 

RU =  7.20 MB/s

LF =   3.54 MB/s

plunger:

plunger run = 801

actual distance =  533.2 um

supposed distance = 532.7 um

micrometer =   533.6 um

position =  300.7 um

voltage =  1.66 V

 

Beam dump: 1-1.5 enA (oscillating)

Colimator :  0.6 enA

 

  63   Fri Mar 3 20:50:31 2023 IK, MP, GJOther Data analysisGe spectra

Ge rates increasing, similar to what it was 24h before. Asking to decrease beam intensity by a factor of 2.

Attached Ge spectra before decreasing.
 

Before decreas beam current was 1.75enA, afterwards 1.1enA.

  159   Wed Mar 8 10:27:26 2023 IK, BSOther GeneralGe spectra

Count rates are still a bit high (~2.6kHz for EAGLE det4). Ge spectra looks fine compared to earlier ones.


 

  164   Wed Mar 8 13:04:05 2023 IKRuns GeneralGe spectra

Ge spectra after #161 - #162 beam tuning.

  206   Fri Mar 10 16:14:19 2023 IPOther GeneralGe - the rate is higher

From 13:00 the rate of the Ge [the chanel 4] is a bit higher (~2.15 kHz). And also the rate of other detectors.

 

  222   Fri Mar 10 22:40:00 2023 IPOther GeneralGe - the rate is higher

the rates of detectors are still a bit higher

Quote:

From 13:00 the rate of the Ge [the chanel 4] is a bit higher (~2.15 kHz). And also the rate of other detectors.

 

 

  135   Tue Mar 7 14:03:20 2023 BSRuns GeneralCounting Rates

 

 

  39   Thu Mar 2 23:04:01 2023 GJProblem Fixed HardwareConnection of the plunger cables, grounding, beam read-out for the operators
Notes of the measurements with different conditions are attached.

Conclusion:
What is read out by the operators on the blue cable should be multiplied by 7 to get a real value in enA.
  230   Sat Mar 11 11:39:55 2023 EGProblems  Buffers clearing

at 11:34:50 sek on 11 march a spike in counting rate of some Ge detectors is observed. Buffers not cleared properly?


 

  62   Fri Mar 3 19:41:27 2023 IKProblems HardwareBeamdump readout issue Vol.2.

Issue described in #61 continues to persist.

Beam stopped from 19:20 to 19:32, run 36 ongoing; MP, GJ, BS entering the experimental hall. 

They changed the position of the grounding point on the chamber, moving it downstream after a fitting/joining part (A to B on the attached picture [artist: GJ]).

 

  61   Fri Mar 3 17:52:41 2023 IKProblem Fixed HardwareBeamdump readout issue - solved

Current readout issues from beamdump. Beam stopped from 17:46-17:52, run 36 ongoing.

MP & GJ checking the grounding and connection the beamdump readout. No visible issue found in the exp hall, but copper grounding wire was tightened, which seemingly resolved the issue (stability of beamdump readout).   

  131   Tue Mar 7 09:56:07 2023 OSTProblems mechanicsBeam stopped and operators controlling number 15_Ge dedector.
  186   Thu Mar 9 13:37:33 2023 OSTOther BeamBeam stopped 13:37-13:49 for control again.
  166   Wed Mar 8 15:57:53 2023 IK, AMProblems GeneralBeam issue

no beam in 15:52 - 16:21 due to power supply problem

  231   Sat Mar 11 12:42:42 2023 EGOther  Beam intensity reduction

Again beam intensity gradually increased reaching 2.4 kHz ontotal rate EAGLE_0 ch4.  Beam intensity reduction requested. 1.8 kHz on ch4 after reduction.

  234   Sat Mar 11 15:16:31 2023 EGOther  Beam intensity reduced again

Again beam intensity to high. Reduced to 2 kHz on ch4 GE

  232   Sat Mar 11 14:15:55 2023 EGRuns  Beam intensity rapid increase

14:12 beam intensity instability. Rapid increase of beam intensity - twice the previous value. Returned to normal after 4 minutes.

  233   Sat Mar 11 14:42:28 2023 EGRuns  Beam instabilities

Large beam intensity instabilities from 14:10 till 14:40.  Beam intensity reduction requested. CH4 on Total rate EAGLE_0 returned to 1.9kHz. Intensity became stable.

 