| ID |
Date |
Author |
Type |
test |
Category |
Subject |
Text |
 |
|
131
|
Tue Mar 7 09:56:07 2023 |
OST | Problems | | mechanics | Beam stopped and operators controlling number 15_Ge dedector. | |
|
|
4
|
Mon Feb 27 12:10:24 2023 |
BS | Other | | Target | target | Support of Target = 1.5 mg/cm2 Ta
Target = 1.0 mg/cm2 Cd-106
Stopper = 10 mg/cm2 Au
|
|
|
92
|
Mon Mar 6 00:30:32 2023 |
MP, ANJ | Other | | Target | new target mounted | New target mounted.
106Cd 1 mg/cm2
181Ta 1.2 mg/cm2 |
|
|
43
|
Fri Mar 3 02:23:45 2023 |
GJ | Problems | | Software | builder gate, TS offsets | Data from current run - 35, with limited beam
current and rates - see a recent post by
Güray.
|
|
|
130
|
Tue Mar 7 09:41:44 2023 |
OST | Problems | | Software | number_Ge_15 dedector not taking data | after run 45 I reset all spectrums and
number_Ge_15 dedector not
taking data. |
|
|
139
|
Tue Mar 7 15:12:06 2023 |
OST | Problems | | Software | run 45 stopped | 14:30 pm to 15:00 run 45 crashed and stopped |
|
|
140
|
Tue Mar 7 15:14:29 2023 |
OST | Problem Fixed | | Software | Run 46 started-15_Ge and number 5 anticompton working now. | VME crate with CAEN digitizers was rebooted.
Run 46 started, number 15_Ge and
number 5 anticompton dedector working |
|
|
223
|
Fri Mar 10 23:27:51 2023 |
IP | Problems | | Software | run 53 - stopped | the daq crashed at 22:43 -- problem with
Local Filter
the run 53 - stopped |
|
|
224
|
Fri Mar 10 23:29:43 2023 |
IP | Problem Fixed | | Software | new run 54 | the daq resolved at 23:15
start new run: run 54 |
|
|
244
|
Sun Mar 12 00:09:35 2023 |
AF | Problem Fixed | | Software | caen_shifted correction in spy & greware | neda_psdcaen_positions.txt had weird values
for three first detectors, especially neda_02.
|
|
|
90
|
Sun Mar 5 16:58:45 2023 |
MP | Problems | | Other | run 37 crashed, run 38 junk | Run 37 stopped by itslelf at 15:43. This
was noticed at about 16:10 by observing that
grafana data rates do not change. |
|
|
195
|
Fri Mar 10 01:28:48 2023 |
AM | Problems | | Other | Plunger distance | longer break during chane of plunger distance
due to technical problem, you can not move
plunger to larger distances, so now we are |
|
|
245
|
Sun Mar 12 03:05:45 2023 |
GJ, AF | Problems | | Network | no data flux visible in grafena, LF prints "write: Broken pipe" | As in the title, see below.
The data are still being constantly written
on the disks with usual rate (also visible |
|
|
39
|
Thu Mar 2 23:04:01 2023 |
GJ | Problem Fixed | | Hardware | Connection of the plunger cables, grounding, beam read-out for the operators | Notes of the measurements with different conditions
are attached.
|
|
|
61
|
Fri Mar 3 17:52:41 2023 |
IK | Problem Fixed | | Hardware | Beamdump readout issue - solved | Current readout issues from beamdump. Beam
stopped from 17:46-17:52, run 36 ongoing.
MP & GJ checking the grounding |
|
|
62
|
Fri Mar 3 19:41:27 2023 |
IK | Problems | | Hardware | Beamdump readout issue Vol.2. | Issue described in #61 continues to
persist.
Beam stopped from 19:20 to 19:32, |
|
|
175
|
Thu Mar 9 01:41:50 2023 |
GJ | Problem Fixed | | Hardware | TS zeroing - cables to S-IN from VME controller - between runs 49 and 50 | Since we were observing jumps of 16 ns in
timings of all channels of whole boards,
the assumption was that the
|
|
|
1
|
Sun Feb 26 16:50:56 2023 |
GJ | Other | | General | proposal | |
|
|
2
|
Sun Feb 26 17:26:45 2023 |
GJ | Other | | General | List of participants | To be filled in, with affiliation and email.
Presumably, for later publication
purposes. |
|
|
3
|
Mon Feb 27 02:49:10 2023 |
GJ | Runs | | General | calibs - 2x 152Eu - run 0 | Run number: 0
Trigger: g
Start time: Mon Feb 27 02:43:50 2023
|
|