EAGLE IRRADIATIONS EXTERNAL Analysis Students Workshop Shift Plans
GAMMA-ONLY NEEDLE NEEDI COULEX PLUNGER GENERAL
HIL102-128Cs HIL087-176Pt Nd148_2 176Pt HIL109-172Os HIL124-128Cs HIL061 Nd148 HIL106-119Cs HIL097-118Xe HIL099-134Sm April16 xdaq
  HIL099-134Sm, Page 8 of 14  ELOG logo
Entry  Fri Mar 3 10:21:18 2023, IP. KWL, Runs, , General, beam intensity back to 1.25 enA at the beam dump 

we increased beam intensity to 1.25 enA (readout at the beam dump)

Entry  Fri Mar 3 13:43:56 2023, IP. KWL, Runs, , , 511 annihilation and 279 Au lines in Ge rings ge_rings101_117_143_acceptedACS_gold_line_279keV.pngge_rings101_117_143_acceptedACS.png

The counting rate in 511 keV line for sum of the Ge detectors placed at 101, 117 and 143 degrees (with approved ACS). The largest intensity is visible for Ge at 143 degrees, lowest for 101 deg.

=============================== RING 101

Integration in spectrum user_ge_ring101_acceptedACS_clone.spc

RESULT : 3895

=============================== RING 117

Integration in spectrum user_ge_ring117_acceptedACS.spc

RESULT : 9392

=============================== RING 143

Integration in spectrum user_ge_ring143_acceptedACS.spc

RESULT : 15767

 

The Coulex of 197Au (279.5 keV ) line is visible in all rings with a comparable intensity:

=============================== RING 101

Integration in spectrum user_ge_ring101_acceptedACS_clone.spc

RESULT : 12491

=============================== RING 117

Integration in spectrum user_ge_ring117_acceptedACS.spc

RESULT : 11443

===============================RING 143

Integration in spectrum user_ge_ring143_acceptedACS.spc

RESULT : 10460

 

 

 

Entry  Fri Mar 3 13:54:41 2023, IP. KWL, Runs, , General, beam energy measurement 

beam stop due to the beam energy measurement

The beam energy: ~145 MeV

Entry  Tue Mar 7 23:39:52 2023, IP, KS, Runs, , General, run 46 stopped 

Run 46 - stopped

  • Run number: 46
  • Start time: Tue Mar 7 15:09:06 2023
  • Stop time: Tue Mar 7 23:38:07 2023
Entry  Tue Mar 7 23:54:05 2023, IP, KS, Runs, , General,  

run control = 47

run data = /onlineNEDA_00/xData/hil099/run_0047_07-03-2023_23h42m03s/needleRU_i47_0047_0000.caendat

RU =  4.47 MB/s

LF =   3.26 MB/s

plunger:

plunger run = 484

actual distance =  282.8 um

supposed distance = 282.7 um

micrometer =   285.2 um

position =  50.1 um

voltage =  2.965 V

 

Beam dump =  0.7 nA

Collimater=  0.5 nA

Entry  Fri Mar 3 11:37:25 2023, IP, Runs, , General,  

plunger run # 230

Beam dump 1.2  enA (given by operators) * 7 (following entry ID#39) = 8.4 enA

Collimator   0.6  enA

Run Data =/onlineNEDA_00/xData/hil099/run_0035_03-03-2023_00h04m26s/needleRU_i35_0035_0047.caendat

RU : 5.88 MB/s

LF : 8.35 MB/s

Supposed Distance = 10 um

position of the target = 10.2 um

micrometer = 8.8 um

Voltage = 7.94 V

Entry  Fri Mar 3 11:48:29 2023, IP, Runs, , General, beam stopped 

The beam is gone.

Entry  Fri Mar 3 12:11:24 2023, IP, Runs, , General,  

the beam comes back.

Entry  Fri Mar 3 14:22:12 2023, IP, Runs, , General,  

the beam comes back. The run 35 is still going.

 

plunger run # 257

Beam dump .... enA (given by operators) * 7 (following entry ID#39) = ... enA

Collimator  ....   enA

Run Data = /onlineNEDA_00/xData/hil099/run_0035_03-03-2023_00h04m26s/needleRU_i35_0035_0059.caendat

RU : 6.18 MB/s

LF : 8.44 MB/s

Supposed Distance = 10 um

position of the target = 10.2 um

micrometer = 8.8 um

Voltage = 7.94 V

 

Entry  Tue Mar 7 17:45:18 2023, IP, Runs, , General,  

run control = 46

run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0013.caendat

RU =  6.10 MB/s

LF =   3.26 MB/s

plunger:

plunger run = 422

actual distance =  282.6um

supposed distance = 282.7 um

micrometer =   284.9 um

position =  49.9 um

voltage =  2.968 V

Beam dump =  0.7 nA

Collimater=  0.5 nA

Entry  Tue Mar 7 19:24:25 2023, IP, Runs, , General,  

run control = 46

run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0022.caendat

RU =  6.22 MB/s

LF =   3.34 MB/s

plunger:

plunger run = 439

actual distance =  282.5um

supposed distance = 282.7 um

micrometer =   284.9 um

position =  49.8 um

voltage =  2.97 V

Beam dump =  0.7 nA

Collimater=  0.5 nA

Entry  Tue Mar 7 20:42:03 2023, IP, Problems, , Beam, beam unstable 

There was a problem with the beam from 20:19 to 20:30.

Now it is fine.

Entry  Tue Mar 7 20:45:13 2023, IP, Runs, , General,  

run control = 46

run data =/onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0028.caendat

RU =  4.9 MB/s

LF =   3.14 MB/s

plunger:

plunger run = 452

actual distance =  282.4 um

supposed distance = 282.7 um

micrometer =   284.7 um

position =  49.7 um

voltage =  2.971 V

Beam dump =  0.7 nA

Collimater=  0.5 nA

Entry  Tue Mar 7 21:55:45 2023, IP, Runs, , General,  

run control = 46

run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0034.caendat

RU =  4.78 MB/s

LF =   3.10 MB/s

plunger:

plunger run = 464

actual distance =  282.8 um

supposed distance = 282.7 um

micrometer =   285.1 um

position =  49.9 um

voltage =  2.965 V

Beam dump =  0.7 nA

Collimater=  0.5 nA

Entry  Thu Mar 9 16:25:57 2023, IP, Runs, , General, run 50 stopped 

run 50 - stopped

  •  
  • Run number: 50
  • Start time: Thu Mar 9 00:50:57 2023
  • Stop time: Thu Mar 9 16:22:37 2023
Entry  Thu Mar 9 19:58:02 2023, IP, Runs, , General,  

run control = 51

run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0018.caendat

RU =  5.93 MB/s

LF =   3.44 MB/s

 

plunger:

plunger run = 927

actual distance =  533.3 um

supposed distance = 532.7 um

micrometer =   533.0 um

position =  300.0 um

voltage =  1.656 V

 

Beam dump: 1 nA

Colimator :  0.6 enA

Entry  Thu Mar 9 21:20:06 2023, IP, Runs, , General,  

run control = 51

run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0026.caendat

RU =  6.23MB/s

LF =   3.56 MB/s

 

plunger:

plunger run = 940

actual distance =  532.2 um

supposed distance = 532.7 um

micrometer =   533.0 um

position =  299.5 um

voltage =  1.659 V

 

Beam dump: 1 nA

Colimator :  0.6 enA

Entry  Thu Mar 9 22:32:01 2023, IP, Runs, , General,  

run control = 51

run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0032.caendat

RU =  6.12 MB/s

LF =   3.42 MB/s

 

plunger:

plunger run = 952

actual distance =  533.2 um

supposed distance = 532.7 um

micrometer =   533.7 um

position =  300.5 um

voltage =  1.656 V

 

Beam dump: 1 nA

Colimator :  0.6 enA

Entry  Fri Mar 10 16:14:19 2023, IP, Other, , General, Ge - the rate is higher Screenshot_2023-03-10_16-10-45.png

From 13:00 the rate of the Ge [the chanel 4] is a bit higher (~2.15 kHz). And also the rate of other detectors.

 

Entry  Fri Mar 10 16:17:57 2023, IP, Runs, , General, run 52 - stopped 

run 52 - stopped

 

  • Run number: 52
  • Start time: Fri Mar 10 01:19:56 2023
  • Stop time: Fri Mar 10 16:16:25 2023
ELOG V3.1.5-fc6679b