Thu Mar 2 09:24:47 2023, ost-mgb, Runs, Run 27 stopped, General,
|
| Run 27 stopped |
Wed Mar 1 19:04:37 2023, BS, Runs, Run, General, Run 26 Started
|
Run 26 Started
Beam 9 enA on the collimator |
Sat Mar 11 10:46:37 2023, EG, Other, Beam intensity, ,
|
ch4 in total rate EGALE_) reached 2.4 kHz, beam intensity has been reduced to fit the required 2kHz.
|
Sun Feb 26 16:50:56 2023, GJ, Other, , General, proposal
|
| |
Sun Feb 26 17:26:45 2023, GJ, Other, , General, List of participants
|
To be filled in, with affiliation and email.
Presumably, for later publication purposes.
|
Mon Feb 27 02:49:10 2023, GJ, Runs, , General, calibs - 2x 152Eu - run 0
|
Run number: 0
Trigger: g
Start time: Mon Feb 27 02:43:50 2023
|
Mon Feb 27 12:10:24 2023, BS, Other, , Target, target
|
Support of Target = 1.5 mg/cm2 Ta
Target = 1.0 mg/cm2 Cd-106
Stopper = 10 mg/cm2 Au
|
Mon Feb 27 12:13:03 2023, MB, Other, , , Plunger Informations
|
Off set of the plunger ~60 um
Distcal file distcal01_vac
|
Mon Feb 27 12:21:29 2023, MP, BS, Runs, , General, List of runs
|
A list of runs, to be filled during the experiment is available here:
https://docs.google.com/spreadsheets/d/11tyI5pqYKJ-rMaYLuBIPZCClWz9Vu87W0R5AZc2srd8/edit?usp=sharing |
Tue Feb 28 23:08:35 2023, BS, Other, , , SHIFT LIST
|
https://docs.google.com/spreadsheets/d/1e2homI9RCYrg369z8aaY3VMIPgapUPOv0zKISsl58vQ/edit?usp=sharing
The data from google sheet obtained 14:17 07/03/23 |
Tue Feb 28 23:10:36 2023, BS, Other, , , What we need to include for Plunger
|
Plunger Run :
Micrometer :
Supposed Distance: |
Wed Mar 1 00:14:28 2023, MP, Other, , Beam, beam energy
|
Beam energy measured: 148.2 MeV
Mind that this may be an overestimate (by ~1-2 MeV) |
Wed Mar 1 11:52:48 2023, MP, Runs, , , first beam
|
| Operators tune beam on the target, collecting run 0023, trigger gg. |
Wed Mar 1 12:51:33 2023, MP, Runs, , Beam, beam on the target
|
Beam on the target:
target+stopper+beam dump: 25 enA (1.7 pnA) not stable
|
Wed Mar 1 16:06:50 2023, BS, Runs, , , Run 23 Stopped
|
Run 23 Stopped due to some unexpected spikes in the
voltage of collimator.
The plunger did not responde to change distance. |
Wed Mar 1 16:07:49 2023, BS, Runs, , , Run 24 Started
|
Run 24 has been started but no beams at the moment.
EAGLE is opened.
Just colecting background. |
Wed Mar 1 20:10:38 2023, BS, DOD, Other, , General,
|
Run 26 is going on but not much for the moment in terms of any useful data.
And half of the EAGLE is opened.
20:50 EAGLE closed. |
Wed Mar 1 21:59:45 2023, MB BS, Other, , General,
|
Offset of the plunger is 60 um.
new distcal01_vac_after_opening |
Thu Mar 2 00:30:29 2023, BS, DOD, Runs, , General,
|
Run 27 with single gamma trigger
Plunger feedback system is not conected.
Data |
Thu Mar 2 04:17:40 2023, DOD, Runs, , General,
|
Run 27 with single gamma trigger
Plunger feedback system is not conected.
Data |
Thu Mar 2 05:33:34 2023, ost-dod, Runs, , General,
|
Run 27 with single gamma trigger
Plunger feedback system is not conected.
Data |
Thu Mar 2 07:40:55 2023, ost-dod, Runs, , General,
|
Run 27 with single gamma trigger
Plunger feedback system is not conected.
Data |
Thu Mar 2 09:05:43 2023, DOD, Problems, , , Sulphur on aluminium fusion-evaporation reactions
|
Using the data which we've sorted to date, I looked into the gamma-gamma coincidences and have identified a number of nuclei e.g. 50Cr, 52Mn,
55Fe which would be consistent with fusion-evaporation reactions involving the sulphur-32 projectiles and aluminium-27. Such reactions would yield a 59Cu
comound system and these channels would be populated as a result of the evaporation of a combination of alphas, protons and neutrons. |
Thu Mar 2 09:20:17 2023, ost-mgb, Runs, , General, Rates
|
Run 27 with single gamma trigger
Plunger feedback system is not conected.
Data |
Thu Mar 2 09:49:13 2023, BS, Runs, , General, run 28 started
|
| run 28 started |
Thu Mar 2 10:18:14 2023, ost-mgb, Runs, , General, run 29 started
|
Run 29 with single gamma trigger
Plunger feedback system is not connected.
Data |
Thu Mar 2 11:14:44 2023, ost-mgb, Runs, , General,
|
un 29 with single gamma trigger
Plunger feedback system is not connected.
Data |
Thu Mar 2 11:17:54 2023, BS, Runs, , , Run 29 stopped
|
| We will change the trigger and we stopped the run number 29 |
Thu Mar 2 11:19:06 2023, BS, Runs, , General, Run 30 Starts
|
We hane now g-g trigger started.
Run 30 started |
Thu Mar 2 12:09:44 2023, ost-mgb, Runs, , General,
|
| beam stop, |
Thu Mar 2 13:50:50 2023, BS, Runs, , General, Run 30 Continueed
|
EAGLE is closed now
We gave a break to test the cable connection of target, degrader, beam dump and collimator.
Run 30 continues to collect data, we did not stop it. |
Thu Mar 2 14:59:32 2023, mgb, Runs, , General,
|
Data |
Thu Mar 2 15:15:40 2023, mgb, Runs, , General, beam stopped. eagle is opened and run 30 continues
|
Data
Position = 522.6
Supposed = 520 |
Thu Mar 2 16:06:17 2023, A.T., Runs, , General,
|
Data
Position = 522.6
Supposed = 520 |
Thu Mar 2 16:40:45 2023, MP, Runs, , General, changing the plunger distance to minimum.
|
Run 31 was stopped, because plunger distance is reduced to the mininimal one.
It takes time, in the meantime, after some break, run 32 was started, with no beam. |
Thu Mar 2 17:23:53 2023, BS, Runs, , General, Plunger Informations
|
Plunger Run : 48
Supposed Distance = 10 um
position of the target = 10 um |
Thu Mar 2 17:25:36 2023, BS, Runs, , General, Run 32 Stopped
|
| Run 32 Stopped |
Thu Mar 2 17:28:21 2023, BS, Runs, , , Run 33 Started
|
Run 33 Started
Plunger Run : 48
Supposed Distance = 10 um |
Thu Mar 2 23:01:21 2023, A.T., Runs, , General, Run 34 started
|
Run 34 Started (33 was stopped due to acquisition problems)
Plunger Run : 105
Supposed Distance = 10 um |
Thu Mar 2 23:04:01 2023, GJ, Problem Fixed, , Hardware, Connection of the plunger cables, grounding, beam read-out for the operators
|
Notes of the measurements with different conditions are attached.
Conclusion:
|
Fri Mar 3 00:09:19 2023, MP, AT, Runs, , , run 35
|
Started after beam current was reduced by a factor of 2
plunger run # 116
Earlier: CUP#6 100 enA |
Fri Mar 3 01:15:27 2023, mgb, Runs, , General,
|
plunger run # 126
Earlier: CUP#6 100 enA |
Fri Mar 3 02:20:46 2023, mgb, Runs, , General,
|
plunger run # 137
Beam dump 2.5 enA |
Fri Mar 3 02:23:45 2023, GJ, Problems, , Software, builder gate, TS offsets
|
Data from current run - 35, with limited beam current and rates - see a recent post by Güray.
BU level, so what we see in Greware (both online & nearline).
Attached is the fold of Germaniums.
|
Fri Mar 3 03:19:34 2023, mgb, Runs, , General,
|
plunger run # 147
Beam dump 2.5 enA |
Fri Mar 3 04:15:32 2023, mgb, Runs, , General,
|
plunger run # 157
Beam dump 2.5 enA |
Fri Mar 3 05:18:35 2023, OST, Runs, , General,
|
plunger run # 167
Beam dump 2.5 enA |
Fri Mar 3 06:24:03 2023, OST, Runs, , General,
|
plunger run # 178
Beam dump 2.5 enA
Collimator 0.9 enA |
Fri Mar 3 07:32:28 2023, OST, Runs, , General,
|
plunger run # 188
Beam dump 2.5 enA
Collimator 0.8 enA |
Fri Mar 3 08:20:13 2023, KWL, Runs, , General,
|
plunger run # 196
Beam dump 1 enA (given by operators) * 7 (following entry ID#39) = 7 enA
Collimator 0.75 enA |
Fri Mar 3 09:30:15 2023, IP. KWL, Runs, , General,
|
plunger run # 208
Beam dump 0.75 enA (given by operators) * 7 (following entry ID#39) = 5.25 enA
Collimator 0.75 enA |
Fri Mar 3 10:09:20 2023, IP. KWL, Runs, , General, beam intensity decreased to 0.3 enA (at beam dump) on our request 
|
// we noticed yesterday that GE at 143 degrees counts more than those placed at 117 and 101. The 511 keV line (+ background) is much more pronounced
for Ge at 143 degrees. To investigate this issue we decreased the beam intensity to 0.3 enA ( value given by operators at the beam dump ).
Conclusion - decrease of the beam intensity does not explain above observations (se attached figs) |
Fri Mar 3 10:21:18 2023, IP. KWL, Runs, , General, beam intensity back to 1.25 enA at the beam dump
|
| we increased beam intensity to 1.25 enA (readout at the beam dump) |
Fri Mar 3 11:37:25 2023, IP, Runs, , General,
|
plunger run # 230
Beam dump 1.2 enA (given by operators) * 7 (following entry ID#39) = 8.4 enA
Collimator 0.6 enA |
Fri Mar 3 11:48:29 2023, IP, Runs, , General, beam stopped
|
| The beam is gone. |
Fri Mar 3 12:11:24 2023, IP, Runs, , General,
|
| the beam comes back. |
Fri Mar 3 13:43:56 2023, IP. KWL, Runs, , , 511 annihilation and 279 Au lines in Ge rings 
|
The counting rate in 511 keV line for sum of the Ge detectors placed at 101, 117 and 143 degrees (with approved ACS). The largest intensity is visible
for Ge at 143 degrees, lowest for 101 deg.
=============================== RING 101 |
Fri Mar 3 13:54:41 2023, IP. KWL, Runs, , General, beam energy measurement
|
beam stop due to the beam energy measurement
The beam energy: ~145 MeV |
Fri Mar 3 14:22:12 2023, IP, Runs, , General,
|
the beam comes back. The run 35 is still going.
plunger run # 257 |
Fri Mar 3 15:04:33 2023, KWL, Other, , General, run 35 stopped
|
| Run 35 stopped. An attempt to check if we can get 5 MeV higher beam energy. |
Fri Mar 3 16:47:29 2023, IK, Runs, , General, Run 36 started
|
Run 36 Started
New Beam energy is 150 MeV
The beam dumb currennt is 2.5 enA. |
Fri Mar 3 17:52:41 2023, IK, Problem Fixed, , Hardware, Beamdump readout issue - solved
|
Current readout issues from beamdump. Beam stopped from 17:46-17:52, run 36 ongoing.
MP & GJ checking the grounding and connection the beamdump readout. No visible issue found in the exp hall, but copper grounding wire
was tightened, which seemingly resolved the issue (stability of beamdump readout). |
Fri Mar 3 19:41:27 2023, IK, Problems, , Hardware, Beamdump readout issue Vol.2.
|
Issue described in #61 continues to persist.
Beam stopped from 19:20 to 19:32, run 36 ongoing; MP, GJ, BS entering the experimental hall.
They changed the position of the grounding point on the chamber, moving it downstream after a fitting/joining part (A to B on the attached |
Fri Mar 3 20:50:31 2023, IK, MP, GJ, Other, , Data analysis, Ge spectra 
|
Ge rates increasing, similar to what it was 24h before. Asking to decrease beam intensity by a factor of 2.
Attached Ge spectra before decreasing.
|
Fri Mar 3 23:38:28 2023, SP, Runs, , General, How different Ge count
|
| All germanium detecors look reasonable, as attached. |
Sat Mar 4 08:27:23 2023, OST, Runs, , General,
|
RUN CONTROL = 36
Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0087.caendat
RU = 4.57 |
Sat Mar 4 09:51:32 2023, OST, Runs, , General,
|
RUN CONTROL = 36
Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0094.caendat
RU = 6.43 |
Sat Mar 4 11:06:31 2023, OST, Runs, , General,
|
RUN CONTROL = 36
Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0100.caendat
RU = 6.03 |
Sat Mar 4 12:34:52 2023, OST, Runs, , General,
|
RUN CONTROL = 36
Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0107.caendat
RU = 6.17 |
Sat Mar 4 13:43:20 2023, OST, Runs, , General,
|
RUN CONTROL = 36
Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0114.caendat
RU = 7.30 |
Sat Mar 4 15:02:54 2023, mgb, Runs, , General, running stable
|
RUN CONTROL = 36
Run Data = /onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0122.caendat
RU = 5.99 |
Sat Mar 4 16:06:19 2023, mgb, Runs, , General,
|
RUN CONTROL = 36
Run Data = /onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0127.caendat
RU = 6.25 |
Sat Mar 4 17:09:13 2023, mgb, Runs, , General,
|
RUN CONTROL = 36
Run Data = /onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0133.caendat
RU = 6.45 MB/s |
Sat Mar 4 18:02:48 2023, mgb, Runs, , General,
|
UN CONTROL = 36
Run Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0138.caendat
RU = 6.39 MB/s |
Sat Mar 4 19:14:30 2023, AF, Runs, , General,
|
AF
UN CONTROL = 36
Run Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0144.caendat |
Sat Mar 4 20:56:40 2023, AF, Runs, , General,
|
AF
UN CONTROL = 36
Run Data =/onlineNEDA_00/xData/hil099/run_0036_03-03-2023_16h47m35s/needleRU_i36_0036_0154.caendat |
Sat Mar 4 23:28:16 2023, AF, Runs, , General,
|
AF
UN CONTROL = 37
Run Data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0000.caendat |
Sun Mar 5 00:08:06 2023, AF, Runs, , General,
|
AF
UN CONTROL = 37
Run Data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0005.caendat |
Sun Mar 5 01:26:58 2023, JD DW, Runs, , General,
|
UN CONTROL = 37
run data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0010.caendat
RU = 6.96 MB/s |
Sun Mar 5 02:06:40 2023, JD DW, Runs, , General,
|
UN CONTROL = 37
run data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0017.caendat
RU = 7.86 MB/s |
Sun Mar 5 03:11:45 2023, JD DW, Runs, , General,
|
RUN CONTROL = 37
run data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0023.caendat
RU = 7.62 MB/s |
Sun Mar 5 04:04:39 2023, JD DW, Runs, , General,
|
UN CONTROL = 37
run data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0030.caendat
RU = 7.78 MB/s |
Sun Mar 5 05:06:50 2023, JD DW, Runs, , General,
|
RUN CONTROL = 37
run data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0037.caendat
RU = 7.93 MB/s |
Sun Mar 5 06:03:26 2023, JD DW, Runs, , General,
|
UN CONTROL = 37
run data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0043.caendat
RU = 7.65 MB/s |
Sun Mar 5 07:12:49 2023, JD DW, Runs, , General,
|
RUN CONTROL = 37
run data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0051.caendat
RU = 8.02 MB/s |
Sun Mar 5 08:54:08 2023, OST, Runs, , General,
|
RUN CONTROL = 37
Data = /onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0061.caendat
RU = 8.12 |
Sun Mar 5 10:17:03 2023, OST, Runs, , General,
|
RUN CONTROL = 37
Data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0072.caendat
RU = 8.01 |
Sun Mar 5 11:45:36 2023, OST, Runs, , General,
|
RUN CONTROL = 37
Data =/onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0082.caendat
RU = 8.50 |
Sun Mar 5 13:52:34 2023, OST, Runs, , General,
|
RUN CONTROL = 37
Data = /onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0097.caendat
RU = 8.17 |
Sun Mar 5 15:07:24 2023, mgb, Runs, , General, Beam dump decreased 1.0 nA to 0.75
|
RUN CONTROL = 37
Data = /onlineNEDA_00/xData/hil099/run_0037_04-03-2023_22h56m17s/needleRU_i37_0037_0106.caendat
RU = 8.65 MB/s |
Sun Mar 5 16:58:45 2023, MP, Problems, , Other, run 37 crashed, run 38 junk
|
Run 37 stopped by itslelf at 15:43. This was noticed at about 16:10 by observing that grafana data rates do not change.
Stop at Run COntrol. LF looks disabled.
Attempt to restart normal way --> run 38 --> failed |
Sun Mar 5 22:32:56 2023, MP, Other, , , change of the target 6x
|
See pictures in the attachment:
1) stopper seen from the back (upstream view)
2) stopper seen from the front (downstream view) |
Mon Mar 6 00:30:32 2023, MP, ANJ, Other, , Target, new target mounted
|
New target mounted.
106Cd 1 mg/cm2
181Ta 1.2 mg/cm2 |
Mon Mar 6 00:43:27 2023, BS, GJ, Runs, , General, Run 40 Started
|
Before this run TSs were moved on the LF level, so the BU has easier job to do and everything is more stable.
BU gate was changed from 1500 to 500 ns, this stay valid from now on.
|
Mon Mar 6 01:00:29 2023, BS, Runs, , General, Run 40 stooped
|
| Run 40 stooped |
Mon Mar 6 01:01:26 2023, BS, Runs, , General, Run 41 started
|
Run 41 started
Plunger distance 10 um |
Mon Mar 6 01:10:59 2023, JD, DW, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0000.caendat
RU = 6.23 MB/s |
Mon Mar 6 02:13:32 2023, JD, DW, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0007.caendat
RU = 7.55 MB/s |
Mon Mar 6 03:17:48 2023, JD, DW, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0014.caendat
RU = 6.94 MB/s |
Mon Mar 6 04:17:48 2023, JD, DW, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0020.caendat
RU = 7.26 MB/s |
Mon Mar 6 05:15:43 2023, JD, DW, Problems, , Beam, No beam 05:10:00
|
| No beam 05:10:00 |
Mon Mar 6 05:16:46 2023, JD, DW, Problem Fixed, , Beam,
|
| Beam back on 05:15 |
Mon Mar 6 05:21:16 2023, JD DW, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0026.caendat
RU = 8.24 MB/s |
Mon Mar 6 06:11:51 2023, JD, DW, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0031.caendat
RU = 7.58 MB/s |
Mon Mar 6 07:19:01 2023, JD, DW, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0038.caendat
RU = 6.46 MB/s |
Mon Mar 6 08:10:32 2023, JD, Runs, , General,
|
RUN CONTROL = 41
Data = /onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0043.caendat
RU = 7.3 MB/s |
Mon Mar 6 09:19:41 2023, OST, Runs, , General,
|
RUN CONTROL = 41
Data =/onlineNEDA_00/xData/hil099/run_0041_06-03-2023_00h58m13s/needleRU_i41_0041_0050.caendat
RU = 8.13 MB/s |
Mon Mar 6 09:53:06 2023, OST, Problems, , Beam, Beam dump decreased 2.5 nA to 1.5 nA- stopped run 41
|
| 9:20 to 9:50 am beam current stayed 2.5 nA . After 9:50 am decrease 1.5 nA |
Mon Mar 6 10:02:47 2023, OST, Problem Fixed, , General, New run started 42
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0000.caendat
/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleLF_i41_0042_0000.adf |
Mon Mar 6 10:56:46 2023, OST, Runs, , General,
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0004.caendat
|
Mon Mar 6 12:10:49 2023, OST, Runs, , General,
|
RUN CONTROL = 42
Data = /onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0010.caendat
RU = 5.46 MB/s |
Mon Mar 6 14:01:48 2023, OST, Runs, , General,
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0019.caendat
RU = 5.12 MB/s |
Mon Mar 6 15:48:23 2023, OST, Runs, , General,
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0026.caendat
RU = 5.24 MB/s |
Mon Mar 6 17:19:24 2023, OST, Problems, , Beam, beam unstable and a little high
|
16:30 beam increased 5 mb/s to 9 mb/s and stayed 1 hour 9 mb/s
17:10 beam fixed. stable |
Mon Mar 6 17:27:42 2023, OST, Runs, , General,
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0035.caendat
RU = 5.43 MB/s |
Mon Mar 6 19:08:51 2023, IK, Runs, , General,
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0045.caendat
RU = 6.95 MB/s |
Mon Mar 6 20:29:45 2023, IK, Runs, , General,
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0053.caendat
RU = 6.75 MB/s |
Mon Mar 6 22:30:46 2023, IK, Runs, , General,
|
RUN CONTROL = 42
Data =/onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0065.caendat
RU = 6.64 MB/s |
Tue Mar 7 00:10:02 2023, JD, DW, Runs, , General,
|
RUN CONTROL = 42
Data = /onlineNEDA_00/xData/hil099/run_0042_06-03-2023_09h54m24s/needleRU_i42_0042_0075.caendat
RU = 6.86 MB/s |
Tue Mar 7 00:54:40 2023, JD, DW, Runs, , General, New run started 43
|
run control = 43
run data = /onlineNEDA_00/xData/hil099/run_0043_07-03-2023_00h50m41s/needleRU_i43_0043_0000.caendat
RU = 7.13 MB/s |
Tue Mar 7 02:02:18 2023, JD, DW, Runs, , General,
|
run control = 43
run data = /onlineNEDA_00/xData/hil099/run_0043_07-03-2023_00h50m41s/needleRU_i43_0043_0006.caendat
RU = 6.27 MB/s |
Tue Mar 7 03:21:33 2023, JD, DW, Problem Fixed, , General, daq crash
|
daq crashed at 02:59:00, resolved at 03:16:00
|
Tue Mar 7 03:26:54 2023, JD, DW, Runs, , General, Run 44 started
|
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0000.caendat
RU = 6.46 MB/s |
Tue Mar 7 04:16:28 2023, JD, DW, Runs, , General,
|
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0005.caendat
RU = 6.49 MB/s |
Tue Mar 7 05:19:55 2023, JD, DW, Runs, , General,
|
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0011.caendat
RU = 6.48 MB/s |
Tue Mar 7 06:18:46 2023, JD, DW, Runs, , General,
|
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0017.caendat
RU =6.95 MB/s |
Tue Mar 7 07:21:04 2023, JD, DW, Runs, , General,
|
run control = 44
run data =/onlineNEDA_00/xData/hil099/run_0044_07-03-2023_03h15m45s/needleRU_i44_0044_0023.caendat
RU = 6.57 MB/s |
Tue Mar 7 09:18:01 2023, OST, Problems, , Beam, run number 44 crashed and local filter turned red-but data saved
|
9:08 local filter turned red. Both .caendat and LF .adf files still incrementing. Stopped run 44, started run 45, looks OK.
|
Tue Mar 7 09:20:53 2023, OST, Problem Fixed, , Beam, run 45 start-local filter turn green
|
| I started run 45, and local filter turn green everything is ok. |
Tue Mar 7 09:24:53 2023, OST, Runs, , General,
|
run control = 45
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0000.caendat
RU = 6.92 MB/s |
Tue Mar 7 09:41:44 2023, OST, Problems, , Software, number_Ge_15 dedector not taking data
|
after run 45 I reset all spectrums and number_Ge_15 dedector not taking data.
i check the fans_monitor_scaler (EAGLE_array_monitor) monitor and number 15 Ge, number ACS 5 never change the color just staying
white there is a problem. |
Tue Mar 7 09:56:07 2023, OST, Problems, , mechanics, Beam stopped and operators controlling number 15_Ge dedector.
|
| |
Tue Mar 7 11:19:54 2023, OST, Runs, , General,
|
run control = 45
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0011.caendat |
Tue Mar 7 12:50:52 2023, OST, Runs, , General,
|
run control = 45
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0019.caendat |
Tue Mar 7 13:51:17 2023, OST, Runs, , General,
|
run control = 45
not fixed number 15_Ge dedector and number 5 anticompton.
run data =/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0025.caendat |
Tue Mar 7 14:03:20 2023, BS, Runs, , General, Counting Rates
|
|
Tue Mar 7 14:07:33 2023, BS, Other, , General, Run Files Number
|
/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleRU_i45_0045_0027.caendat
/onlineNEDA_00/xData/hil099/run_0045_07-03-2023_09h14m17s/needleLF_i44_0045_0010.adf |
Tue Mar 7 14:23:43 2023, BS, Other, , General, RU and BU Rates
|
| |
Tue Mar 7 14:39:08 2023, BS, Other, , General, Runlist File ScreenShot Accesses on 07 MArch at 14:38
|
| |
Tue Mar 7 15:12:06 2023, OST, Problems, , Software, run 45 stopped
|
| 14:30 pm to 15:00 run 45 crashed and stopped |
Tue Mar 7 15:14:29 2023, OST, Problem Fixed, , Software, Run 46 started-15_Ge and number 5 anticompton working now.
|
VME crate with CAEN digitizers was rebooted.
Run 46 started, number 15_Ge and number 5 anticompton dedector working now.
|
Tue Mar 7 16:06:16 2023, OST, Runs, , General,
|
run control = 46
run data =/onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0004.caendat
RU = 6.51 MB/s |
Tue Mar 7 17:45:18 2023, IP, Runs, , General,
|
run control = 46
run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0013.caendat
RU = 6.10 MB/s |
Tue Mar 7 19:24:25 2023, IP, Runs, , General,
|
run control = 46
run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0022.caendat
RU = 6.22 MB/s |
Tue Mar 7 20:42:03 2023, IP, Problems, , Beam, beam unstable
|
There was a problem with the beam from 20:19 to 20:30.
Now it is fine. |
Tue Mar 7 20:45:13 2023, IP, Runs, , General,
|
run control = 46
run data =/onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0028.caendat
RU = 4.9 MB/s |
Tue Mar 7 21:55:45 2023, IP, Runs, , General,
|
run control = 46
run data = /onlineNEDA_00/xData/hil099/run_0046_07-03-2023_15h08m38s/needleRU_i46_0046_0034.caendat
RU = 4.78 MB/s |
Tue Mar 7 23:39:52 2023, IP, KS, Runs, , General, run 46 stopped
|
Run 46 - stopped
Run number: 46
Start time: Tue Mar 7 15:09:06 2023
Stop time: Tue Mar 7 23:38:07 2023
|
Tue Mar 7 23:54:05 2023, IP, KS, Runs, , General,
|
run control = 47
run data = /onlineNEDA_00/xData/hil099/run_0047_07-03-2023_23h42m03s/needleRU_i47_0047_0000.caendat
RU = 4.47 MB/s |
Wed Mar 8 00:31:05 2023, KS, KSz, Runs, , General, Run 47 standard info
|
run control = 47
run data = /onlineNEDA_00/xData/hil099/run_0047_07-03-2023_23h42m03s/needleRU_i47_0047_0003.caendat
RU = 5.12 MB/s |
Wed Mar 8 01:24:04 2023, BS, Runs, , General, Run 47 Stopped to Change Distance Change
|
Run 47 stopped and distance changed from 50 um (+ 100 um offset) to 150 um (+ 100 um offset)
Last plunger run for 50 um (+ 100 um offset) is 495 |
Wed Mar 8 01:24:58 2023, KS, KSz, Runs, , General, New run started 48 (new distance)
|
run control = 48
run data = /onlineNEDA_00/xData/hil099/run_0048_08-03-2023_01h08m57s/needleRU_i48_0048_0000.caendat
RU = 5.09 MB/s |
Wed Mar 8 02:45:53 2023, KS, KSz, Other, , Beam, increase in Counting Rates
|
Since 2:32 increased counting rate observed:
Sinde 3:45 decrease to 1.5 kHz ch4.
run control = 48 |
Wed Mar 8 04:24:40 2023, KS, KSz, Runs, , General, Run 48
|
run control = 48
run data = /onlineNEDA_00/xData/hil099/run_0048_08-03-2023_01h08m57s/needleRU_i48_0048_0017.caendat
RU = 5.75 MB/s |
Wed Mar 8 06:25:53 2023, KS, KSz, Runs, , General,
|
run control = 48
run data = /onlineNEDA_00/xData/hil099/run_0048_08-03-2023_01h08m57s/needleRU_i48_0048_0029.caendat
RU = 6.59 MB/s |
Wed Mar 8 07:30:15 2023, KS, KSz, Problems, , Beam, beam unstable
|
| The beam was unstable from 7:18 to 7:25. |
Wed Mar 8 07:57:05 2023, IK, Problems, , Beam, beam unstable
|
The beam was unstable from 7:55. Stabilized around 8:03.
Counting rates are higher, EAGLE det4 went to ~2.5kHZ from ~2kHz (attached).
According to the operator, a bit of a tuning was done, current is 1nA. |
Wed Mar 8 09:31:43 2023, IK, Runs, , General, Run 48 stopped, run 49 started
|
Run 48 stopped at 9:29 to start a new run.
Run 49 is started okay at 9:31.
|
Wed Mar 8 10:27:26 2023, IK, BS, Other, , General, Ge spectra
|
Count rates are still a bit high (~2.6kHz for EAGLE det4). Ge spectra looks fine compared to earlier ones.
|
Wed Mar 8 10:31:10 2023, IK, Runs, , General, Run 49
|
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0008.caendat
RU = 10.6 MB/s |
Wed Mar 8 11:42:14 2023, IK, Problems, , General, Run 49  
|
Counting rates continued to increase (det4 > 3kHz). Beam current looked fine on the beam dump, but beam is fluctuating, and can reach 2nA on the collimator.
Asked for improved focusing.
Attached screenshots and Ge spectra from before changing anything. |
Wed Mar 8 12:11:59 2023, IK, Problem Fixed, , General, beam tuned
|
Beam tuning was made, focusing significantly improved.
Intensity: |
Wed Mar 8 13:02:38 2023, IK, Runs, , General, Run 49
|
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0030.caendat
RU = 7.86 MB/s |
Wed Mar 8 13:04:05 2023, IK, Runs, , General, Ge spectra
|
| Ge spectra after #161 - #162 beam tuning. |
Wed Mar 8 14:47:34 2023, AM, Runs, , Beam, beam unstable
|
| two drops in beam intesitivity at 14:33, 14:37 and 14:51 |
Wed Mar 8 15:57:53 2023, IK, AM, Problems, , General, Beam issue
|
| no beam in 15:52 - 16:21 due to power supply problem |
Wed Mar 8 16:51:13 2023, BS, Runs, , General,
|
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0050.caendat
RU = 6.75 MB/s |
Wed Mar 8 16:53:07 2023, BS, Runs, , General, Rates 
|
| |
Wed Mar 8 17:59:38 2023, BS, Runs, , General, General
|
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0058.caendat
RU = 9.83 MB/s |
Wed Mar 8 18:02:43 2023, BS, Runs, , General, Rates
|
| |
Wed Mar 8 19:01:49 2023, OST, Runs, , General, beam current was very low and beam increased after this entry
|
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0064.caendat
RU = 5.75 MB/s |
Wed Mar 8 20:46:16 2023, OST, Runs, , General,
|
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0074.caendat
RU = 6.35 MB/s |
Wed Mar 8 22:36:30 2023, OST, Runs, , General,
|
run control = 49
run data = /onlineNEDA_00/xData/hil099/run_0049_08-03-2023_09h31m08s/needleRU_i49_0049_0082.caendat
RU = 6.06 MB/s |
Thu Mar 9 01:07:48 2023, WP, Runs, , General, New run started 50 (distance change)
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0000.caendat
|
Thu Mar 9 01:41:50 2023, GJ, Problem Fixed, , Hardware, TS zeroing - cables to S-IN from VME controller - between runs 49 and 50
|
Since we were observing jumps of 16 ns in timings of all channels of whole boards, the assumption was that the
signals from logical FIFO arrive at different times to S-IN inputs to zero the TS counters.
6 originally placed cables of quite the same length were replaced with 6 cables of declared 3-ns length, same
|
Thu Mar 9 02:14:09 2023, WP, Runs, , General, General
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0008.caendat
|
Thu Mar 9 03:14:39 2023, WP, Runs, , General, General
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0015.caendat
|
Thu Mar 9 04:17:21 2023, WP, Runs, , General, General
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0022.caendat
|
Thu Mar 9 05:19:33 2023, WP, Runs, , General, General
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0028.caendat
|
Thu Mar 9 06:24:10 2023, WP, Runs, , General, General
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0035.caendat
|
Thu Mar 9 07:28:59 2023, WP, Runs, , General, General
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0042.caendat
|
Thu Mar 9 08:54:13 2023, OST, Runs, , General,
|
run control = 50
run data =/onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0051.caendat
RU = 6.86 MB/s |
Thu Mar 9 10:05:30 2023, OST, Runs, , General,
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0058.caendat
RU = 7.48 MB/s |
Thu Mar 9 12:06:30 2023, OST, Runs, , General,
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0070.caendat
RU = 6.89 MB/s |
Thu Mar 9 13:22:37 2023, OST, Other, , Beam, beam stopped 13:21-13:24 for control
|
| |
Thu Mar 9 13:37:33 2023, OST, Other, , Beam, Beam stopped 13:37-13:49 for control again.
|
| |
Thu Mar 9 15:17:32 2023, OST, Runs, , General,
|
run control = 50
run data = /onlineNEDA_00/xData/hil099/run_0050_09-03-2023_00h50m28s/needleRU_i50_0050_0087.caendat
RU = 6.04 MB/s |
Thu Mar 9 16:25:57 2023, IP, Runs, , General, run 50 stopped
|
run 50 - stopped
Run number: 50
Start time: Thu Mar 9 00:50:57 2023
Stop time: Thu Mar 9 16:22:37 2023
|
Thu Mar 9 16:54:27 2023, MK, Runs, , General,
|
run control = 51
run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0002.caendat
RU = 6.43MB/s |
Thu Mar 9 17:57:41 2023, MW, Runs, , General,
|
run control = 51
run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0007.caendat
RU = 6.45MB/s |
Thu Mar 9 19:58:02 2023, IP, Runs, , General,
|
run control = 51
run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0018.caendat
RU = 5.93 MB/s |
Thu Mar 9 21:20:06 2023, IP, Runs, , General,
|
run control = 51
run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0026.caendat
RU = 6.23MB/s |
Thu Mar 9 22:32:01 2023, IP, Runs, , General,
|
run control = 51
run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0032.caendat
RU = 6.12 MB/s |
Fri Mar 10 00:09:32 2023, AM, Runs, , General,
|
run control = 51
run data = /onlineNEDA_00/xData/hil099/run_0051_09-03-2023_16h25m16s/needleRU_i51_0051_0040.caendat
RU = 5.56 MB/s |
Fri Mar 10 01:28:48 2023, AM, Problems, , Other, Plunger distance
|
longer break during chane of plunger distance due to technical problem, you can not move plunger to larger distances, so now we are choosing smaller
distance in the morning we ask the specialists
|
Fri Mar 10 01:32:20 2023, AM, Runs, , General, run 51 stopped
|
run 51 stopped
Run number: 51
Start time: Thu Mar 9 16:25:00 2023
Stop time: Fri Mar 10 00:13:21 2023
|
Fri Mar 10 03:20:42 2023, AM, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0010.caendat
RU = 5.55 MB/s |
Fri Mar 10 05:40:34 2023, AM, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0022.caendat
RU = 6.59 MB/s |
Fri Mar 10 07:03:54 2023, AM, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0029.caendat
RU = 6.06 MB/s |
Fri Mar 10 08:02:39 2023, AM, Problems, , Beam, bigger data rate
|
rate of ch4 above 2kHz, too big beam intesivity, highier data rate from: 07:51
|
Fri Mar 10 10:02:30 2023, OST, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0045.caendat
RU = 6.18 MB/s |
Fri Mar 10 10:03:36 2023, OST, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0045.caendat
RU = 6.18 MB/s |
Fri Mar 10 11:36:23 2023, OST, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0053.caendat
RU = 5.69 MB/s |
Fri Mar 10 13:35:49 2023, OST, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0064.caendat
RU = 7.06 MB/s |
Fri Mar 10 15:04:16 2023, OST, Runs, , General,
|
run control = 52
run data = /onlineNEDA_00/xData/hil099/run_0052_10-03-2023_01h19m26s/needleRU_i52_0052_0074.caendat
RU = 7.17 MB/s |
Fri Mar 10 16:14:19 2023, IP, Other, , General, Ge - the rate is higher
|
From 13:00 the rate of the Ge [the chanel 4] is a bit higher (~2.15 kHz). And also the rate of other detectors.
|
Fri Mar 10 16:17:57 2023, IP, Runs, , General, run 52 - stopped
|
run 52 - stopped
Run number: 52
Start time: Fri Mar 10 01:19:56 2023
Stop time: Fri Mar 10 16:16:25 2023
|
Fri Mar 10 16:23:21 2023, IP, Runs, , General,
|
run control = 53
run data = /onlineNEDA_00/xData/hil099/run_0053_10-03-2023_16h18m36s/needleRU_i53_0053_0000.caendat |
Fri Mar 10 17:34:07 2023, IP, Runs, , General,
|
run control = 53
run data = /onlineNEDA_00/xData/hil099/run_0053_10-03-2023_16h18m36s/needleRU_i53_0053_0007.caendat |
Fri Mar 10 18:42:02 2023, IP, Runs, , General,
|
run control = 53
run data = /onlineNEDA_00/xData/hil099/run_0053_10-03-2023_16h18m36s/needleRU_i53_0053_0014.caendat |
Fri Mar 10 19:30:48 2023, IP, Runs, , Beam, stop beam
|
the beam was stopped at 19:27
|
Fri Mar 10 19:32:01 2023, IP, Runs, , Beam, start beam
|
| the beam start: 19:31 |
Fri Mar 10 19:42:57 2023, IP, Runs, , General,
|
run control = 53
run data =/onlineNEDA_00/xData/hil099/run_0053_10-03-2023_16h18m36s/needleRU_i53_0053_0020.caendat |
Fri Mar 10 20:54:20 2023, IP, Problems, , Beam, lost the beam
|
| the beam was lost at 20:46. The operator is working to solve the problem. |
Fri Mar 10 21:00:15 2023, IP, Problem Fixed, , Beam, the beam comes back
|
| the beam comes back at 20:54 |
Fri Mar 10 21:15:01 2023, IP, Runs, , General,
|
run control = 53
run data =/onlineNEDA_00/xData/hil099/run_0053_10-03-2023_16h18m36s/needleRU_i53_0053_0029.caendat |
Fri Mar 10 21:24:29 2023, IP, Runs, , Beam, the beam stopped
|
| the beam is stopped at 21:23. |
Fri Mar 10 21:26:55 2023, IP, Problems, , Beam, problem with focus the beam
|
| the beam is not the best focused. The operator works on this problem. |
Fri Mar 10 21:29:28 2023, IP, Runs, , Beam, beam start
|
the beam starts at 21:27.
But the operators are still working on focusing the beam. |
Fri Mar 10 21:30:42 2023, IP, Problem Fixed, , Beam, the beam is focused
|
| the beam is now good focused |
Fri Mar 10 22:36:55 2023, IP, Runs, , General,
|
run control = 53
run data = /onlineNEDA_00/xData/hil099/run_0053_10-03-2023_16h18m36s/needleRU_i53_0053_0037.caendat |
Fri Mar 10 22:40:00 2023, IP, Other, , General, Ge - the rate is higher
|
the rates of detectors are still a bit higher
Quote:
From 13:00 the rate of the Ge [the chanel 4] is a bit higher (~2.15 kHz). And also the rate of other detectors. |
Fri Mar 10 23:27:51 2023, IP, Problems, , Software, run 53 - stopped
|
the daq crashed at 22:43 -- problem with Local Filter
the run 53 - stopped |
Fri Mar 10 23:29:43 2023, IP, Problem Fixed, , Software, new run 54
|
the daq resolved at 23:15
start new run: run 54 |
Fri Mar 10 23:35:28 2023, IP, Runs, , General,
|
run control = 54
run data = run_0054_10-03-2023_23h14m24s/needleRU_i54_0054_0001.caendat |
Sat Mar 11 00:05:06 2023, IP, Runs, , General, run 54 - stopped
|
run 54 - stopped due to change the distance in plunger
Run number: 54
Start time: Fri Mar 10 23:14:54 2023
Stop time: Sat Mar 11 00:02:37 2023
|
Sat Mar 11 00:34:51 2023, BS, Runs, , General, Run 55 Started New Distance 500 um
|
| Run 55 Started New Distance 500 um |
Sat Mar 11 09:29:48 2023, EG, Other, , , current exeriment parameters as requested in the manual for shifters
|
41_ge_single_theta 101,117,143 looks ok. No peakdeformation of 511, other gammas FWHM around half of that of511 - ok
ch4 in total rate EAGLE_0 arouns 2.1 kHz - ok
planger tollerace limits (green and blue horizontal markers) set far of the current distance, intantionally? |
Sat Mar 11 11:39:55 2023, EG, Problems, , , Buffers clearing
|
at 11:34:50 sek on 11 march a spike in counting rate of some Ge detectors is observed. Buffers not cleared properly?
|
Sat Mar 11 12:42:42 2023, EG, Other, , , Beam intensity reduction
|
| Again beam intensity gradually increased reaching 2.4 kHz ontotal rate EAGLE_0 ch4. Beam intensity reduction requested. 1.8 kHz on ch4 after reduction. |
Sat Mar 11 14:15:55 2023, EG, Runs, , , Beam intensity rapid increase
|
| 14:12 beam intensity instability. Rapid increase of beam intensity - twice the previous value. Returned to normal after 4 minutes. |
Sat Mar 11 14:42:28 2023, EG, Runs, , , Beam instabilities
|
Large beam intensity instabilities from 14:10 till 14:40. Beam intensity reduction requested. CH4 on Total rate EAGLE_0 returned to 1.9kHz. Intensity
became stable.
|
Sat Mar 11 15:16:31 2023, EG, Other, , , Beam intensity reduced again
|
| Again beam intensity to high. Reduced to 2 kHz on ch4 GE |
Sat Mar 11 16:22:06 2023, OST, Runs, , General,
|
run control = 55
run data = /onlineNEDA_00/xData/hil099/run_0055_11-03-2023_00h33m38s/needleRU_i55_0055_0103.caendat
RU = 8.23 MB/s |
Sat Mar 11 17:58:35 2023, OST, Runs, , Beam,
|
| beam stopped 17:54. |
Sat Mar 11 18:06:36 2023, OST, Runs, , Beam,
|
| beam started 18:04. |
Sat Mar 11 18:10:06 2023, OST, Runs, , General,
|
run control = 55
run data = /onlineNEDA_00/xData/hil099/run_0055_11-03-2023_00h33m38s/needleRU_i55_0055_0114.caendat
RU = 5.72 MB/s |
Sat Mar 11 19:14:35 2023, BS, Runs, , General, Rates 
|
Run files /onlineNEDA_00/xData/hil099/run_0055_11-03-2023_00h33m38s/needleRU_i55_0055_0114.caendat
Beam Dump = 0.7 nA
Run number 55 |
Sat Mar 11 21:24:04 2023, OST, Runs, , General,
|
run control = 55
run data = /onlineNEDA_00/xData/hil099/run_0055_11-03-2023_00h33m38s/needleRU_i55_0055_0129.caendat
RU = 6.45 MB/s |
Sat Mar 11 22:27:08 2023, BS, Runs, , General, Rates
|
Rates 22:18 367Hz
Rates 22:22 4.78 kHz
And Now
see the figure
|
Sat Mar 11 22:52:48 2023, BS, Runs, , General, run_0055
|
run control = 55
run data = /onlineNEDA_00/xData/hil099/run_0055_11-03-2023_00h33m38s/needleRU_i55_0055_0137.caendat
RU = 4.17 MB/s |
Sat Mar 11 23:30:47 2023, AF, Runs, , General, run_0056
|
run control = 56
run data =/onlineNEDA_00/xData/hil099/run_0056_11-03-2023_23h26m10s
RU = 7.3 MB/s |
Sun Mar 12 00:09:35 2023, AF, Problem Fixed, , Software, caen_shifted correction in spy & greware
|
neda_psdcaen_positions.txt had weird values for three first detectors, especially neda_02.
It was:
|
Sun Mar 12 03:05:45 2023, GJ, AF, Problems, , Network, no data flux visible in grafena, LF prints "write: Broken pipe"
|
As in the title, see below.
The data are still being constantly written on the disks with usual rate (also visible on grafena).
Greware is also seeing data.
|
Sun Mar 12 03:23:05 2023, AF, Runs, , General, run_0057
|
run control = 57
run data =/onlineNEDA_00/xData/hil099/run_0057_12-03-2023_03h10m10s/
RU = 8.1 MB/s |
Sun Mar 12 04:20:19 2023, OST, Runs, , General, Run 57 stopped, Run 58 started
|
| |
Sun Mar 12 07:01:31 2023, OST, Runs, , General,
|
run control = 58
run data =xdaq 2254808224 Mar 12 06:59 needleRU_i58_0058_0017.caendat
RU = 8.10 MB/s |
Sun Mar 12 08:03:08 2023, OST, Runs, , General,
|
run control = 58
run data = xdaq 3303602776 Mar 12 08:01 needleRU_i58_0058_0024.caendat
RU = 7.88 MB/s |
Sun Mar 12 09:15:00 2023, EG, Other, , , Array_monitors mat refresh synchronization
|
Beam stable, counting rate within limits, plunger distance within the given tollerance, gamma spectra ok, everything looks good but....
(yes there must be some "but"), refreshing of EAGLE_array_monitor.mat and NEDA_array_monitor.mat are out of sync. Mostly pure red colour
is dysplayed. Not very usefull and (I suppose) missleading information. |
Sun Mar 12 13:20:00 2023, EG, Problems, , , No Data
|
For a split second a red "NO DATA" notification appeared in the top right corner of Counting rates dashboard.
It looks like it did not change anything in data collection.
|
Sun Mar 12 17:03:12 2023, OST, Runs, , General,
|
run control = 58
run data =xdaq 1111949504 Mar 12 17:01 needleRU_i58_0058_0083.caendat
RU = 7.14 MB/s |
Sun Mar 12 18:50:37 2023, OST, Runs, , General,
|
run control = 58
run data =xdaq 471656980 Mar 12 18:47 needleRU_i58_0058_0095.caendat
RU = 7.35 MB/s |
Sun Mar 12 19:00:59 2023, OST, Runs, , General, Run 58 stopped and Run 59 started.
|
18:54 Run 58 stopped and 18:55 Run 59 started.
BEAM STOPPED.
END of EXPERIMENT. |
Sun Mar 12 19:20:37 2023, MP, Runs, , General, run 60, radioactivity,
|
Run 59 collected data using gg trigger radioactivty
Run 60 trigger changed fropm gg to single g.
Just in the beginning a few minutes gg data could have been collected. |
Sun Mar 12 20:31:45 2023, GJ, Runs, , General, run_61 - target and stopper and beam dump dismounted.
|
target and stopper and beam dump dismounted.
only the constructional elements of the plunger remained.
|
Sun Mar 12 20:55:08 2023, MP, Runs, , , runc62 152 for efficiency calibration
|
started
source in the target position, as precisely as possible
|
Sun Mar 12 21:33:54 2023, GJ, Runs, , General, run 63 - 152 Eu during LN2 filling
|
| |
Sun Mar 12 22:52:36 2023, Bahadir Saygi, Other, , General, Plunger Run Files 
|
| Plunger run files for later to calculate error in the distance |
Mon Mar 13 12:15:35 2023, MP, Other, , , measurement of radioactivity along the beam line
|
At about 12:00, 13/03/2023 (about 17 hours after the beam was stopped), a measurement of activity of the beam line was performed using FH
40G-L10 meter.
Radioctivity run 65 stopped before the measurement, radioactivity run 66 started after the measurement |
Mon Mar 13 12:16:31 2023, MP, Other, , , run 66
|
run 66 started, radiocativity (background)
|
Mon Mar 13 15:16:21 2023, MP, Other, , , list of files copied to Bahadir's disks 
|
Attached are the lists of files copied to the two 4 TB Bahadir's disks.
|
Thu Mar 16 12:13:11 2023, MP, Other, , , test of radioactivity  
|
16/03/2023
Plunger chamber closed, some dummy target and stopper probably inside
run 72 eagle closed |
Mon Mar 20 04:55:24 2023, MP, Other, , , missing caendat.bz2 on Bahadir's disk
|
The following caendat.bz2 files were not copied to Bahadir's disks
(because they were not bzipped at the time of the copying):
|
Tue Jan 23 16:55:21 2024, yes, Other, , , 106Cd certificate
|
| |
|